A 6µW ECG-Recording ΔΣ Modulator with Internal-Capacitor-Flipping Technique for 34V<sub>pp</sub> Common-Mode-Interference (CMI) Tolerance and 1V<sub>pp</sub> Input Range

A two-electrode ECG system is widely used in wearable devices for continuous cardiac and cardiovascular health status monitoring, but several challenges must be addressed for reliable signal acquisition. One of the key challenges is the 50/60Hz common-mode interference (CMI) of several volts or more, caused by coupling between the body and powerline, resulting in saturation of the ECG-recording front-end circuit. To address this, ECG-recording ICs tolerant to a large CMI have been actively researched [1]–[7]. Two representative techniques are CM charge compensation (CMCC) [1]–[6] and CM averaging (CMA) [7] (Fig. 1 (top)). In the CMCC approach, CMI can be cancelled through charge compensation using switched capacitors [1]–[3] or current sources [4]–[6]. However, implementing such an additional compensation circuit and its controller incurs additional area consumption and significant power overhead. The work reported in [3] proposes a circuit implementing CMCC without a controller to minimize the circuit overhead, but compensation circuit overhead still remains. Moreover, the CMCC circuit adds extra noise to the ECG-recording system, degrading the noise performance [1]. In the CMA approach, the whole ECG-recording circuit ground is driven by the detected CMI to nullify the effect of CMI. However, its CMI tolerance performance is sensitive to the parasitic capacitance <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">$(\mathrm{C}_{\mathrm{P}}]$</tex> present between the circuit ground and earth ground. Since Cp varies significantly as the surrounding environment changes, very large averaging capacitors (C<inf xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">AVG</inf>) and high-driving-capability buffers are needed to provide sufficient CMI tolerance even for the worst-case C<inf xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">P</inf>, resulting in large area and power overhead. Another key challenge is motion artifacts (MAs), which are caused by user movement and contain significant differential-mode (DM) components. A wide input range is required for securing tolerance against MAs. Importantly, all these prior designs addressing the CMI challenge [1]–[7] are based on the conventional ECG-recording circuit structure consisting of an instrumentation amplifier (IA) and a SAR ADC. This structure inherently cannot support a wide input range, making them susceptible to MAs. In [6], the MA problem is addressed by tracking and precharging DM artifacts, but this incurs a large power overhead.

Authors

Institutions

Publication Details

Published
2025-04-13
DOI
https://doi.org/10.1109/cicc63670.2025.10982871
Primary Topic
ECG Monitoring and Analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A 6µW ECG-Recording ΔΣ Modulator with Internal-Capacitor-Flipping Technique for 34Vpp Common-Mode-Interference (CMI) Tolerance and 1Vpp Input Range

Minkyu Je, Sohmyung Ha, Vincent Lukito, Jimin Koo et al.
ECG Monitoring and Analysis
article

A 6µW ECG-Recording ΔΣ Modulator with Internal-Capacitor-Flipping Technique for 34Vpp Common-Mode-Interference (CMI) Tolerance and 1Vpp Input Range

Minkyu Je, Sohmyung Ha, Vincent Lukito, Jimin Koo, Sein Oh, Yoontae Jung
article en

Abstract

A two-electrode ECG system is widely used in wearable devices for continuous cardiac and cardiovascular health status monitoring, but several challenges must be addressed for reliable signal acquisition. One of the key challenges is the 50/60Hz common-mode interference (CMI) of several volts or more, caused by coupling between the body and powerline, resulting in saturation of the ECG-recording front-end circuit. To address this, ECG-recording ICs tolerant to a large CMI have been actively researched [1]–[7]. Two representative techniques are CM charge compensation (CMCC) [1]–[6] and CM averaging (CMA) [7] (Fig. 1 (top)). In the CMCC approach, CMI can be cancelled through charge compensation using switched capacitors [1]–[3] or current sources [4]–[6]. However, implementing such an additional compensation circuit and its controller incurs additional area consumption and significant power overhead. The work reported in [3] proposes a circuit implementing CMCC without a controller to minimize the circuit overhead, but compensation circuit overhead still remains. Moreover, the CMCC circuit adds extra noise to the ECG-recording system, degrading the noise performance [1]. In the CMA approach, the whole ECG-recording circuit ground is driven by the detected CMI to nullify the effect of CMI. However, its CMI tolerance performance is sensitive to the parasitic capacitance <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">$(\mathrm{C}_{\mathrm{P}}]$</tex> present between the circuit ground and earth ground. Since Cp varies significantly as the surrounding environment changes, very large averaging capacitors (C<inf xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">AVG</inf>) and high-driving-capability buffers are needed to provide sufficient CMI tolerance even for the worst-case C<inf xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">P</inf>, resulting in large area and power overhead. Another key challenge is motion artifacts (MAs), which are caused by user movement and contain significant differential-mode (DM) components. A wide input range is required for securing tolerance against MAs. Importantly, all these prior designs addressing the CMI challenge [1]–[7] are based on the conventional ECG-recording circuit structure consisting of an instrumentation amplifier (IA) and a SAR ADC. This structure inherently cannot support a wide input range, making them susceptible to MAs. In [6], the MA problem is addressed by tracking and precharging DM artifacts, but this incurs a large power overhead.

New York University Abu Dhabi (AE), Korea Advanced Institute of Science and Technology (KR), IMEC (BE)
Affordable and clean energy
Openalex Percentile: Top 81%
ECG Monitoring and Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.