Angular dependence of detector responses in high-energy brachytherapy

<i>Objective.</i>To characterize three small volume ionization chambers and a diamond detector concerning their angular dependent response under Ir-192 irradiation.<i>Approach.</i>Monte Carlo (MC) simulations as well as experimental measurements of the angular dependent response were performed at 10 cm distance in air. Further simulations under TG-43 conditions were performed at distances of 1, 3, 5 and 10 cm in water. The detectors were placed in relation to their reference points specified by the manufacturer and rotated relative to the beam axis. Results for the different setups were compared to investigate the influence of water and air as well as a distance dependence of the angular response. Detector specific properties such as dead volumes of ionization chambers or asymmetries were taken into account.<i>Main results.</i>The MC simulations in air could be experimentally verified. The detectors showed an angular dependence<0.5%for rotations⩽20∘,⩽35∘,⩽40∘and⩽50∘at distances of 1, 3, 5 and 10 cm in water, respectively. In air at 10 cm distance, the angular dependence exceeded 0.5% for rotations of 20<sup>∘</sup>for the microDiamond and for 50<sup>∘</sup>for the Semiflex and PinPoint. Here, the angular dependence of the A26 was<0.5%at all angles. The slope of the angular dependence increased with rotation from frontal alignment for all detectors. Variations of the detector angular dependent response caused by volume averaging were largest close to the source, i.e. up to 8% at 1 cm distance. High-Z materials in the vicinity of the sensitive volumes were an additional source of angular response variations amounting up to 4% independent of the distance.<i>Significance.</i>To the best of the authors' knowledge, this work is the first investigation of perturbation factors, also differential in energy, leading to angular dependent detector response in high-energy brachytherapy dosimetry. The results could be useful to estimate the angular dependent response when measuring complex dose rate distributions, for example multiple source positions or in a heterogeneous environment.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-05-19
DOI
https://doi.org/10.1088/1361-6560/adda8c
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Angular dependence of detector responses in high-energy brachytherapy

Klemens Zink, Thomas Failing, T. Schneider, Boris Keil et al.
Physics in Medicine and Biology
Advanced Radiotherapy Techniques
article

Angular dependence of detector responses in high-energy brachytherapy

Klemens Zink, Thomas Failing, T. Schneider, Boris Keil, R. Behrens, Frank W. Hensley
article en

Abstract

<i>Objective.</i>To characterize three small volume ionization chambers and a diamond detector concerning their angular dependent response under Ir-192 irradiation.<i>Approach.</i>Monte Carlo (MC) simulations as well as experimental measurements of the angular dependent response were performed at 10 cm distance in air. Further simulations under TG-43 conditions were performed at distances of 1, 3, 5 and 10 cm in water. The detectors were placed in relation to their reference points specified by the manufacturer and rotated relative to the beam axis. Results for the different setups were compared to investigate the influence of water and air as well as a distance dependence of the angular response. Detector specific properties such as dead volumes of ionization chambers or asymmetries were taken into account.<i>Main results.</i>The MC simulations in air could be experimentally verified. The detectors showed an angular dependence<0.5%for rotations⩽20∘,⩽35∘,⩽40∘and⩽50∘at distances of 1, 3, 5 and 10 cm in water, respectively. In air at 10 cm distance, the angular dependence exceeded 0.5% for rotations of 20<sup>∘</sup>for the microDiamond and for 50<sup>∘</sup>for the Semiflex and PinPoint. Here, the angular dependence of the A26 was<0.5%at all angles. The slope of the angular dependence increased with rotation from frontal alignment for all detectors. Variations of the detector angular dependent response caused by volume averaging were largest close to the source, i.e. up to 8% at 1 cm distance. High-Z materials in the vicinity of the sensitive volumes were an additional source of angular response variations amounting up to 4% independent of the distance.<i>Significance.</i>To the best of the authors' knowledge, this work is the first investigation of perturbation factors, also differential in energy, leading to angular dependent detector response in high-energy brachytherapy dosimetry. The results could be useful to estimate the angular dependent response when measuring complex dose rate distributions, for example multiple source positions or in a heterogeneous environment.

Physics in Medicine and BiologyVol. 70(11)
Physikalisch-Technische Bundesanstalt (DE), Philipps University of Marburg (DE), Heidelberg University (DE), University Hospital Heidelberg (DE), Heidelberger Institut für Radioonkologie (DE), Technische Hochschule Mittelhessen (DE)
Affordable and clean energy
Openalex Percentile: Top 87%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.