Impact of implant abutment materials on force damping response and marginal fit of implant supported restoration
Lithium disilicate exhibited the smallest marginal gap before and after Cementation, while PEEK showed the largest, followed by Titanium and resin-ceramic material. Resin-ceramic material had the highest shock absorption for force damping, followed by PEEK and Lithium disilicate, while Titanium recorded the highest impact force, indicating the least damping ability.
Authors
- Amany Korsel
- Waleed Elshahawy
- Amr Mohammed Hussein Elmesery
Institutions
- Tanta University (EG)
Publication Details
- Journal
- BMC Oral Health
- Published
- 2025-05-19
- DOI
- https://doi.org/10.1186/s12903-025-06112-0
- Primary Topic
- Dental Implant Techniques and Outcomes
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Tanta University