Integrated pipeline for minimally invasive extracorporeal circulation: A basic experimental study

ObjectiveThis study aims to evaluate the operational feasibility of a self-designed minimally invasive extracorporeal circulation integrated circuit (miECC) type IV system and to assess the effectiveness of the venous gas filter and automatic exhaust mechanism.MethodsA blood recycler was utilized in place of a heart to connect the arteriovenous line, venous gas filter, centrifugal pump head, and integrated membrane-lung connection, thereby establishing a loop circulation model for the miECC IV system. The venous gas filter was linked to a negative pressure suction device positioned at its top. During operation, gas was introduced from the venous end to determine the maximum gas storage capacity of the venous gas filter, the gas removal efficiency of the negative pressure suction device, the conversion capabilities between open and closed operations, and the adequacy of the pipeline design.ResultsThe minimally invasive extracorporeal circulation integrated pipeline demonstrated stable operational performance and successfully facilitated the transition between closed and open states. The venous gas filter exhibited a gas storage capacity of less than 80 mL across various flow rates. Experimental results indicated that at flow rates of 3.0-4.5 L/min, the automatic venting device effectively removed gas when the volume did not exceed 80 mL. At flow rates of 5.0-5.5 L/min, the device also successfully vented gas under the same conditions.ConclusionThe self-designed minimally invasive extracorporeal circulation integrated pipeline operates effectively and can transition between closed and open configurations. The gas filter effectively prevents venous gas accumulation. At the same time, the enhanced automatic gas-exhausting device, utilizing negative pressure, efficiently removes gas from the venous filter, thereby improving the safety of the closed-circulation operation.

Authors

Institutions

Publication Details

Journal
Perfusion
Published
2025-05-19
DOI
https://doi.org/10.1177/02676591251340976
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Integrated pipeline for minimally invasive extracorporeal circulation: A basic experimental study

Yuheng Lang, Yanying Wang, Zhiyong Dong, Chengxiu Zhao et al.
Perfusion
Mechanical Circulatory Support Devices
article

Integrated pipeline for minimally invasive extracorporeal circulation: A basic experimental study

Yuheng Lang, Yanying Wang, Zhiyong Dong, Chengxiu Zhao, Haiping Guo, Rui Li, Haiping Liu, Shuang Gao, Jian Yang, Liang Li
article en

Abstract

ObjectiveThis study aims to evaluate the operational feasibility of a self-designed minimally invasive extracorporeal circulation integrated circuit (miECC) type IV system and to assess the effectiveness of the venous gas filter and automatic exhaust mechanism.MethodsA blood recycler was utilized in place of a heart to connect the arteriovenous line, venous gas filter, centrifugal pump head, and integrated membrane-lung connection, thereby establishing a loop circulation model for the miECC IV system. The venous gas filter was linked to a negative pressure suction device positioned at its top. During operation, gas was introduced from the venous end to determine the maximum gas storage capacity of the venous gas filter, the gas removal efficiency of the negative pressure suction device, the conversion capabilities between open and closed operations, and the adequacy of the pipeline design.ResultsThe minimally invasive extracorporeal circulation integrated pipeline demonstrated stable operational performance and successfully facilitated the transition between closed and open states. The venous gas filter exhibited a gas storage capacity of less than 80 mL across various flow rates. Experimental results indicated that at flow rates of 3.0-4.5 L/min, the automatic venting device effectively removed gas when the volume did not exceed 80 mL. At flow rates of 5.0-5.5 L/min, the device also successfully vented gas under the same conditions.ConclusionThe self-designed minimally invasive extracorporeal circulation integrated pipeline operates effectively and can transition between closed and open configurations. The gas filter effectively prevents venous gas accumulation. At the same time, the enhanced automatic gas-exhausting device, utilizing negative pressure, efficiently removes gas from the venous filter, thereby improving the safety of the closed-circulation operation.

PerfusionVol. 41(1)
Handan College (CN), Tianjin University of Technology (CN), Tianjin University (CN), Tianjin Third Central Hospital (CN), Membrane Technology & Research (United States) (US), Tianjin Hospital (CN), Tianjin Medical University (CN)
Affordable and clean energy
Openalex Percentile: Top 90%
Mechanical Circulatory Support Devices
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.