Retained bioburden does not pose contamination risk after a full sterile processing cycle

Introduction After sterile processing, orthopedic instrument trays are occasionally found to have retained bone from prior procedures. This study aims to assess the effect on heavily contaminated bioburden, adjacent tray, and instrument contamination risk of a standard medical facility sterile processing cycle for orthopedic instruments. Methods Porcine bone was placed in a wire mesh basket and inoculated with Geobacillus Stearotherophilus (G. stearothermophilus). This was then processed in the standard manner at our medical center, placed into unwrapped trays, and then into a multi-tray sterilization container (Turbett Sterilization Pod) for autoclave sterilization. The study was conducted at a private laboratory specializing in certification of sterile processing, Highpower ValidationTesting and Lab Services, Rochester ,N.Y.(Highpower).The steps of processing included an initial soak, ultrasonic cleaning, rinse, automated wash cycle, and steam sterilization. Following this, all test samples were placed into culture medium and incubated for seven days. Results There were no positive test samples following completion of the standard sterile processing steps. These included 36 samples of inoculated bone, 72 samples of simulated adjacent instruments, and 36 spore indicator strips. Conclusions This is the first investigation to the authors knowledge that examines bone bioburden in the context of a standard full sterile processing sequence. Similar prior studies have focused on the autoclave only efficacy. This study suggests that bone bioburden that has been through the full sequence of sterile processing is not contaminated and does not cross contaminate adjacent instruments or trays. The methodology used most closely replicated the scenario of freely exposed bone bioburden.

Authors

Institutions

Publication Details

Journal
Journal of Orthopaedic Experience & Innovation
Published
2025-05-18
DOI
https://doi.org/10.60118/001c.133635
Primary Topic
Medical Device Sterilization and Disinfection
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Retained bioburden does not pose contamination risk after a full sterile processing cycle

Sridhar R. Rachala, Mary T. Bayers-Thering, Brian J. McGrath, Matthew J. Phillips et al.
Journal of Orthopaedic Experience & Innovation
Medical Device Sterilization and Disinfection
article

Retained bioburden does not pose contamination risk after a full sterile processing cycle

Sridhar R. Rachala, Mary T. Bayers-Thering, Brian J. McGrath, Matthew J. Phillips, Aaron Leininger, Andrew Stegemann
article en

Abstract

Introduction After sterile processing, orthopedic instrument trays are occasionally found to have retained bone from prior procedures. This study aims to assess the effect on heavily contaminated bioburden, adjacent tray, and instrument contamination risk of a standard medical facility sterile processing cycle for orthopedic instruments. Methods Porcine bone was placed in a wire mesh basket and inoculated with Geobacillus Stearotherophilus (G. stearothermophilus). This was then processed in the standard manner at our medical center, placed into unwrapped trays, and then into a multi-tray sterilization container (Turbett Sterilization Pod) for autoclave sterilization. The study was conducted at a private laboratory specializing in certification of sterile processing, Highpower ValidationTesting and Lab Services, Rochester ,N.Y.(Highpower).The steps of processing included an initial soak, ultrasonic cleaning, rinse, automated wash cycle, and steam sterilization. Following this, all test samples were placed into culture medium and incubated for seven days. Results There were no positive test samples following completion of the standard sterile processing steps. These included 36 samples of inoculated bone, 72 samples of simulated adjacent instruments, and 36 spore indicator strips. Conclusions This is the first investigation to the authors knowledge that examines bone bioburden in the context of a standard full sterile processing sequence. Similar prior studies have focused on the autoclave only efficacy. This study suggests that bone bioburden that has been through the full sequence of sterile processing is not contaminated and does not cross contaminate adjacent instruments or trays. The methodology used most closely replicated the scenario of freely exposed bone bioburden.

Journal of Orthopaedic Experience & InnovationVol. 6(1)
Bucknell University (US), Yale University (US), University at Buffalo, State University of New York (US)
Openalex Percentile: Top 90%
Medical Device Sterilization and Disinfection
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.