The effect to dose and diagnostic quality of limiting source-to-image distance on anterior-posterior semi-erect adult chest X-rays

The findings demonstrate that in controlled X-ray settings, the ESD is increased when the SID is reduced from 180cm to 150cm SID (40.1%) and to 100cm SID (121.5%), which furthermore decreases image quality (p = 0.03) and the diagnostic value of the image. It is recommended that radiographers maintain the standard 180 cm SID during adult inpatient semi-erect AP CXR examinations to maintain as low as reasonably practicable (ALARP) diagnostic X-ray examinations.

Authors

Institutions

Publication Details

Journal
Journal of medical imaging and radiation sciences
Published
2025-05-17
DOI
https://doi.org/10.1016/j.jmir.2025.101986
Citations
1
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
1.36
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The effect to dose and diagnostic quality of limiting source-to-image distance on anterior-posterior semi-erect adult chest X-rays

P. Lockwood, Anita Fitzgerald
1 citations
Journal of medical imaging and radiation sciences
Radiation Dose and Imaging
1.36
article

The effect to dose and diagnostic quality of limiting source-to-image distance on anterior-posterior semi-erect adult chest X-rays

P. Lockwood, Anita Fitzgerald
article en
1 citations

Abstract

The findings demonstrate that in controlled X-ray settings, the ESD is increased when the SID is reduced from 180cm to 150cm SID (40.1%) and to 100cm SID (121.5%), which furthermore decreases image quality (p = 0.03) and the diagnostic value of the image. It is recommended that radiographers maintain the standard 180 cm SID during adult inpatient semi-erect AP CXR examinations to maintain as low as reasonably practicable (ALARP) diagnostic X-ray examinations.

Journal of medical imaging and radiation sciencesVol. 56(4)
Canterbury Christ Church University (GB)
Gender equality
Openalex Percentile: Top 20%
Radiation Dose and Imaging
1.36
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

The effect to dose and diagnostic quality of limiting source-to-image distance on anterior-posterior semi-erect adult chest X-rays — P. Lockwood, Anita Fitzgerald · Journal of medical imaging and radiation sciences (2025) | TGRS Research Map | TGRS