A glucometer-colorimetric dual mode biosensor for point-of-care testing of cardiac troponin I

cTnI is considered the primary marker for the diagnosis of Acute myocardial infarction (AMI).. At present, cTnI relies on laboratory testing, which requires technical personnel and specialized equipment. This strategy proposes a dual-mode sensor based on a blood glucose meter and colorimetry to achieve portable, rapid, and accurate measurement of cTnI. The feasibility of this sensor was validated using UV spectrophotometry and polyacrylamide gel electrophoresis. Under optimal conditions, the linear ranges for the glucometer and colorimetric modes were 0.01 ngmL -1 to 20.0 ngmL -1 and 0.03 ngmL -1 to 15.0 ngmL -1 , with detection limits of 0.001 ngmL -1 and 0.005 ngmL -1 , respectively. The dual-mode sensor exhibits high sensitivity and selectivity, enabling rapid portable dual-mode detection. This method has been successfully applied for detecting cTnI in human serum, indicating significant potential for clinical applications.

Authors

Institutions

Publication Details

Journal
Arabian Journal of Chemistry
Published
2025-05-15
DOI
https://doi.org/10.25259/ajc_178_2024
Primary Topic
Advanced biosensing and bioanalysis techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A glucometer-colorimetric dual mode biosensor for point-of-care testing of cardiac troponin I

喜三郎 花沢, Ruomu Chen, Lingmei Niu, Jingyi Wang et al.
Arabian Journal of Chemistry
Advanced biosensing and bioanalysis techniques
article

A glucometer-colorimetric dual mode biosensor for point-of-care testing of cardiac troponin I

喜三郎 花沢, Ruomu Chen, Lingmei Niu, Jingyi Wang, Wei Yuan, Kejia Xu, Meng Jiang, Yan Jin, Wen Yan, Huajie Li, Kai Kang
article en

Abstract

cTnI is considered the primary marker for the diagnosis of Acute myocardial infarction (AMI).. At present, cTnI relies on laboratory testing, which requires technical personnel and specialized equipment. This strategy proposes a dual-mode sensor based on a blood glucose meter and colorimetry to achieve portable, rapid, and accurate measurement of cTnI. The feasibility of this sensor was validated using UV spectrophotometry and polyacrylamide gel electrophoresis. Under optimal conditions, the linear ranges for the glucometer and colorimetric modes were 0.01 ngmL -1 to 20.0 ngmL -1 and 0.03 ngmL -1 to 15.0 ngmL -1 , with detection limits of 0.001 ngmL -1 and 0.005 ngmL -1 , respectively. The dual-mode sensor exhibits high sensitivity and selectivity, enabling rapid portable dual-mode detection. This method has been successfully applied for detecting cTnI in human serum, indicating significant potential for clinical applications.

Arabian Journal of ChemistryVol. 18
Hebei Medical University (CN)
Openalex Percentile: Top 90%
Advanced biosensing and bioanalysis techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.