<b><i>Objectives:</i></b> The aim of this study is to investigate the impact of water mist flow rate on the energy stability of Erbium - doped Yttrium Aluminum Garnet (Er:YAG) laser and ablation of bone. <b><i>Materials and Methods:</i></b> The pig tibiae surface was ablated with Er:YAG laser (100 mJ, 20 Hz, 100 μs) at the flow rate of water mist of 0, 3, 6, 9, 12 mL/min, respectively, for 1 min each. An energy meter was used to record the energy of each laser pulse under different water mist flow rates. A thermocouple was inserted into the bone marrow cavity to record the temperature changes. Infrared thermography, stereomicroscope, and electronic balance were employed to record the surface temperature, morphology, and mass changes, respectively. <b><i>Results:</i></b> As the water mist flow rate escalates, the laser energy experiences a notable attenuation of 27.4%, and the stability parameter Root Mean Square (RMS) escalates by a factor of 30. Despite the relatively minor temperature change within the bone marrow cavity, obviously different surface morphologies are evident. In cases where water mist is less than 3 mL/min, the surface temperature soars, resulting in carbonization. Conversely, when the water mist is greater than 9 mL/min, the ablating mass deteriorates significantly, to merely 25.2% of the value without water mist cooling. <b><i>Conclusions:</i></b> The water mist can affect the laser energy reaching the bone surface and its stability. When the water mist is insufficient, surface temperatures rise, causing carbonization. Conversely, it significantly reduces the ablating mass. Therefore, it is crucial to adjust the amount of water mist appropriately in erbium laser ablated bone. Given such large fluctuations in laser energy, the accuracy of erbium laser bone ablation is dramatically reduced, and the use of water mist cooling should be carefully considered.

Authors

Institutions

Publication Details

Journal
Photobiomodulation Photomedicine and Laser Surgery
Published
2025-05-15
DOI
https://doi.org/10.1089/photob.2025.0028
Primary Topic
Laser Applications in Dentistry and Medicine
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effects of Water Mist Flow Rate in Erbium - Yttrium Aluminum Garnet Laser Ablated Bone

Ziran Liu, Shengjie Ma, Shanshan Tu, Ke Li et al.
Photobiomodulation Photomedicine and Laser Surgery
Laser Applications in Dentistry and Medicine
article

Effects of Water Mist Flow Rate in Erbium - Yttrium Aluminum Garnet Laser Ablated Bone

Ziran Liu, Shengjie Ma, Shanshan Tu, Ke Li, Jiliang Xu, Zhengdong Xiong, Guanyu Zhang, Hao Wu
article en

Abstract

<b><i>Objectives:</i></b> The aim of this study is to investigate the impact of water mist flow rate on the energy stability of Erbium - doped Yttrium Aluminum Garnet (Er:YAG) laser and ablation of bone. <b><i>Materials and Methods:</i></b> The pig tibiae surface was ablated with Er:YAG laser (100 mJ, 20 Hz, 100 μs) at the flow rate of water mist of 0, 3, 6, 9, 12 mL/min, respectively, for 1 min each. An energy meter was used to record the energy of each laser pulse under different water mist flow rates. A thermocouple was inserted into the bone marrow cavity to record the temperature changes. Infrared thermography, stereomicroscope, and electronic balance were employed to record the surface temperature, morphology, and mass changes, respectively. <b><i>Results:</i></b> As the water mist flow rate escalates, the laser energy experiences a notable attenuation of 27.4%, and the stability parameter Root Mean Square (RMS) escalates by a factor of 30. Despite the relatively minor temperature change within the bone marrow cavity, obviously different surface morphologies are evident. In cases where water mist is less than 3 mL/min, the surface temperature soars, resulting in carbonization. Conversely, when the water mist is greater than 9 mL/min, the ablating mass deteriorates significantly, to merely 25.2% of the value without water mist cooling. <b><i>Conclusions:</i></b> The water mist can affect the laser energy reaching the bone surface and its stability. When the water mist is insufficient, surface temperatures rise, causing carbonization. Conversely, it significantly reduces the ablating mass. Therefore, it is crucial to adjust the amount of water mist appropriately in erbium laser ablated bone. Given such large fluctuations in laser energy, the accuracy of erbium laser bone ablation is dramatically reduced, and the use of water mist cooling should be carefully considered.

Photobiomodulation Photomedicine and Laser SurgeryVol. 43(6)
Anhui Medical University (CN), Chinese Academy of Sciences (CN), Anhui University of Traditional Chinese Medicine (CN), Hefei Institutes of Physical Science (CN), Chinese PLA General Hospital (CN), Anhui Institute of Optics and Fine Mechanics (CN), Second Affiliated Hospital of Anhui Medical University (CN)
Clean water and sanitation
Openalex Percentile: Top 84%
Laser Applications in Dentistry and Medicine
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.