Evaluation of the performance of three pulse oximeters at different probe positions in awake rabbits

In rabbits, pulse oximetry plays a crucial role in monitoring oxygen saturation (SpO2) and pulse rate (PR), but it can be challenging due to their small size, variable anatomy, limited measurement sites, and potential motion artifacts during measurements. This prospective randomized study aimed to assess the reliability of three pulse oximeters on the forelimb and tail of fifty awake, healthy European rabbits. Two pulse oximeters, the Masimo Rad-5 (device 1) and the Edan VE-H100B, were tested, with the latter using both a Y-clip (device 2) and the Nellcor OxiMax adhesive finger sensor (device 3). Reliable values of SpO2 and PR were considered ≥ 95% and ≥ 151 bpm, respectively, which are normal values in a healthy rabbit. Success rates of the devices were calculated as the ratio of reliable to unreliable/missing values and compared using χ2 tests. Correlation and agreement between PR and clinically measured heart rate (cHR) were assessed using the Pearson correlation coefficient (r) and Bland-Altman analysis, respectively. Success rates differed significantly among devices (p < 0.001). The SpO2 success rates were 46%, 95.6%, and 94.4% for devices 1, 2, and 3, respectively. The PR success rates were 54.6%, 63%, and 72.8% for devices 1, 2, and 3, respectively. Success rates were higher on the forelimb (>66%) than on the tail (>26%) across devices. Correlation between PR and cHR was stronger on forelimb (r > 0.80) than on tail (r < 0.70). Agreement between PR and cHR was similar among devices, with a mean bias ranging from -6.6 to 6.7 bpm, and limits of agreement from -33.8 to 41.5 bpm. The Edan devices showed superior ability to detect reliable values of SpO2 and PR compared to the Masimo in rabbits. The forelimb appears to be a more reliable site for pulse oximetry in rabbits than the tail.

Authors

Institutions

Publication Details

Journal
PLoS ONE
Published
2025-05-15
DOI
https://doi.org/10.1371/journal.pone.0323044
Citations
3
Primary Topic
Veterinary Pharmacology and Anesthesia
Type
article
Field-Weighted Citation Impact
7.06
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of the performance of three pulse oximeters at different probe positions in awake rabbits

Barbara Contiero, Vinícius Rosa, Martina Cardinali, Francesca Zanusso et al.
3 citations
PLoS ONE
Veterinary Pharmacology and Anesthesia
7.06
article

Evaluation of the performance of three pulse oximeters at different probe positions in awake rabbits

Barbara Contiero, Vinícius Rosa, Martina Cardinali, Francesca Zanusso, Giulia Maria De Benedictis, Daniele Bovo
article en
3 citations

Abstract

In rabbits, pulse oximetry plays a crucial role in monitoring oxygen saturation (SpO2) and pulse rate (PR), but it can be challenging due to their small size, variable anatomy, limited measurement sites, and potential motion artifacts during measurements. This prospective randomized study aimed to assess the reliability of three pulse oximeters on the forelimb and tail of fifty awake, healthy European rabbits. Two pulse oximeters, the Masimo Rad-5 (device 1) and the Edan VE-H100B, were tested, with the latter using both a Y-clip (device 2) and the Nellcor OxiMax adhesive finger sensor (device 3). Reliable values of SpO2 and PR were considered ≥ 95% and ≥ 151 bpm, respectively, which are normal values in a healthy rabbit. Success rates of the devices were calculated as the ratio of reliable to unreliable/missing values and compared using χ2 tests. Correlation and agreement between PR and clinically measured heart rate (cHR) were assessed using the Pearson correlation coefficient (r) and Bland-Altman analysis, respectively. Success rates differed significantly among devices (p < 0.001). The SpO2 success rates were 46%, 95.6%, and 94.4% for devices 1, 2, and 3, respectively. The PR success rates were 54.6%, 63%, and 72.8% for devices 1, 2, and 3, respectively. Success rates were higher on the forelimb (>66%) than on the tail (>26%) across devices. Correlation between PR and cHR was stronger on forelimb (r > 0.80) than on tail (r < 0.70). Agreement between PR and cHR was similar among devices, with a mean bias ranging from -6.6 to 6.7 bpm, and limits of agreement from -33.8 to 41.5 bpm. The Edan devices showed superior ability to detect reliable values of SpO2 and PR compared to the Masimo in rabbits. The forelimb appears to be a more reliable site for pulse oximetry in rabbits than the tail.

PLoS ONEVol. 20(5)
University of Padua (IT), Southern Oncology Specialists (US)
Openalex Percentile: Top 3%
Veterinary Pharmacology and Anesthesia
7.06
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.