Comparison of Three Point-of-Care Analyzers for the Measurement of Lactate Concentration in Chelonians

Abstract Blood lactate concentration has been identified as a prognostic indicator in chelonians. However, comparison among various studies about lactate concentration is hindered by the fact that various point-of-care analyzers are used. This study evaluated agreement, correlation and biases between three commonly used analyzers, the Lactate Plus (Nova), i-STAT (Abbott), and EPOC (Heska). Forty chelonians of 10 different species were included in the study. The animals were recruited from a rehabilitation center and from the clientele of a veterinary teaching hospital, and presented with a variety of health statuses and a wide range of lactate concentration. A single blood sample obtained from each individual was placed in a heparinized tube and analyzed concomitantly by two analyzers per sample (20 chelonians for each analyzer). Agreement between the lactate and blood analyzers was evaluated via a Bland–Altman plot and correlation via a Spearman correlation coefficient. Bias was determined using a Passing–Bablok regression analysis. There was a fair agreement between the three techniques, but agreement was decreased for lactate concentrations above 5 mmol/L. Both blood analyzers reported consistently greater lactate concentrations compared to values obtained with the Lactate Plus, with a positive proportional bias. Correlations between the Lactate Plus and the two other point-of-care analyzers were excellent, 97.0% and 92.8% with the i-STAT and the EPOC analyzers, respectively. This study highlights the need to conduct studies in various reptile species to evaluate the performance of point-of-care analyzers.

Authors

Institutions

Publication Details

Journal
Journal of Herpetological Medicine and Surgery
Published
2025-01-01
DOI
https://doi.org/10.5818/jhms-d-24-00038
Citations
1
Primary Topic
Advanced Chemical Sensor Technologies
Type
article
Field-Weighted Citation Impact
0.67
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of Three Point-of-Care Analyzers for the Measurement of Lactate Concentration in Chelonians

Claire Vergneau‐Grosset, Édouard Maccolini, Alex Boutin
1 citations
Journal of Herpetological Medicine and Surgery
Advanced Chemical Sensor Technologies
0.67
article

Comparison of Three Point-of-Care Analyzers for the Measurement of Lactate Concentration in Chelonians

Claire Vergneau‐Grosset, Édouard Maccolini, Alex Boutin
article en
1 citations

Abstract

Abstract Blood lactate concentration has been identified as a prognostic indicator in chelonians. However, comparison among various studies about lactate concentration is hindered by the fact that various point-of-care analyzers are used. This study evaluated agreement, correlation and biases between three commonly used analyzers, the Lactate Plus (Nova), i-STAT (Abbott), and EPOC (Heska). Forty chelonians of 10 different species were included in the study. The animals were recruited from a rehabilitation center and from the clientele of a veterinary teaching hospital, and presented with a variety of health statuses and a wide range of lactate concentration. A single blood sample obtained from each individual was placed in a heparinized tube and analyzed concomitantly by two analyzers per sample (20 chelonians for each analyzer). Agreement between the lactate and blood analyzers was evaluated via a Bland–Altman plot and correlation via a Spearman correlation coefficient. Bias was determined using a Passing–Bablok regression analysis. There was a fair agreement between the three techniques, but agreement was decreased for lactate concentrations above 5 mmol/L. Both blood analyzers reported consistently greater lactate concentrations compared to values obtained with the Lactate Plus, with a positive proportional bias. Correlations between the Lactate Plus and the two other point-of-care analyzers were excellent, 97.0% and 92.8% with the i-STAT and the EPOC analyzers, respectively. This study highlights the need to conduct studies in various reptile species to evaluate the performance of point-of-care analyzers.

Journal of Herpetological Medicine and SurgeryVol. 35(2)
Cegep de Saint Hyacinthe (CA), Ministère des Ressources naturelles et des Forêts (Québec) (CA), Université Laval (CA), Université de Montréal (CA)
Openalex Percentile: Top 32%
Advanced Chemical Sensor Technologies
0.67
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.