Establishment of a Radiation Dose–Response Calibration Curve Using a Rapid Cytokinesis-block Micronucleus Assay for Dose Assessment of Medical Radiation Equipment

Dicentric chromosome analysis (DCA) has limitations in its use for the evaluation of the radiation dose upon the development of medical radiation equipment due to its time/labour-consuming procedure and the requirement of highly trained experts. Therefore, we aimed to construct a dose-response curve using a semi-automatic cytokinesis-block micronucleus (CBMN) analysis method that can be easily analysed and utilised by anyone. CHO-K1 cells were exposed to gamma rays at various doses (0-4 Gy). For the CBMN assay, the bi-nucleated cells were selected and captured, and micronuclei (MN) scoring was automatically performed using the Metafer4 system. The MN scores were manually confirmed and corrected by analysts. Using the frequency distributions of MN according to the radiation dose, the dose-response calibration curve was generated using Dose Estimate v5.2 software. The equation of dose-response calibration curve is Y = 0.0299 (±0.0057) + 0.1502 (±0.0151) × D + 0.0111 (±0.0048) × D<sup>2</sup>. The goodness-of-fit parameters were also calculated (chi-squared [<i>χ</i> <sup>2</sup>] = 39.45, degrees of freedom = 5, <i>p</i> = 0.0000). The semi-automated CBMN assay consist of two steps: the automated MN capture/scoring step and the manual confirmation/correction step. Using an established dose-response calibration curve and the procedure of the semi-automated CBMN assay, the dose-estimation of gamma-irradiated (0.5 or 2 Gy) CHO-K1 cells were performed by two analysts individually, and it was inter-compared to verify the accuracy, the results showed that the estimated doses were a good fit the applied doses of radiation. The CBMN assay using CHO-K1 cells can be easily used as a biodosimetry tool for dose assessment of medical radiation equipment due to the advantage of being simple, easy, and quick to measure the dose.

Authors

Institutions

Publication Details

Journal
Genome Integrity
Published
2025-01-01
DOI
https://doi.org/10.14293/genint.15.1.003
Primary Topic
Carcinogens and Genotoxicity Assessment
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Establishment of a Radiation Dose–Response Calibration Curve Using a Rapid Cytokinesis-block Micronucleus Assay for Dose Assessment of Medical Radiation Equipment

Il Hyeok Seo, Mooyoung Yoo, Se Hyun Kim, S.H. Cho et al.
Genome Integrity
Carcinogens and Genotoxicity Assessment
article

Establishment of a Radiation Dose–Response Calibration Curve Using a Rapid Cytokinesis-block Micronucleus Assay for Dose Assessment of Medical Radiation Equipment

Il Hyeok Seo, Mooyoung Yoo, Se Hyun Kim, S.H. Cho, Ji Young Kim, Hoon Sik Choi, Jeong-In Kim
article en

Abstract

Dicentric chromosome analysis (DCA) has limitations in its use for the evaluation of the radiation dose upon the development of medical radiation equipment due to its time/labour-consuming procedure and the requirement of highly trained experts. Therefore, we aimed to construct a dose-response curve using a semi-automatic cytokinesis-block micronucleus (CBMN) analysis method that can be easily analysed and utilised by anyone. CHO-K1 cells were exposed to gamma rays at various doses (0-4 Gy). For the CBMN assay, the bi-nucleated cells were selected and captured, and micronuclei (MN) scoring was automatically performed using the Metafer4 system. The MN scores were manually confirmed and corrected by analysts. Using the frequency distributions of MN according to the radiation dose, the dose-response calibration curve was generated using Dose Estimate v5.2 software. The equation of dose-response calibration curve is Y = 0.0299 (±0.0057) + 0.1502 (±0.0151) × D + 0.0111 (±0.0048) × D<sup>2</sup>. The goodness-of-fit parameters were also calculated (chi-squared [<i>χ</i> <sup>2</sup>] = 39.45, degrees of freedom = 5, <i>p</i> = 0.0000). The semi-automated CBMN assay consist of two steps: the automated MN capture/scoring step and the manual confirmation/correction step. Using an established dose-response calibration curve and the procedure of the semi-automated CBMN assay, the dose-estimation of gamma-irradiated (0.5 or 2 Gy) CHO-K1 cells were performed by two analysts individually, and it was inter-compared to verify the accuracy, the results showed that the estimated doses were a good fit the applied doses of radiation. The CBMN assay using CHO-K1 cells can be easily used as a biodosimetry tool for dose assessment of medical radiation equipment due to the advantage of being simple, easy, and quick to measure the dose.

Genome IntegrityVol. 16
Korea Electric Power Corporation (South Korea) (KR), Daewon Pharm (South Korea) (KR)
Decent work and economic growth
Openalex Percentile: Top 86%
Carcinogens and Genotoxicity Assessment
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.