Concentration Estimation of Apixaban and Rivaroxaban in Plasma From Patients Through Heparin‐Calibrated Anti‐Xa Assay Generated Different Results by Using 3 Reagent/Calibrators/Instrument Systems
Estimation of APIXA or RIVA concentration through HC-Anti Xa is possible, but values obtained are different for each reagent/calibrator/instrument system. Knowledge of the HC-Anti Xa values corresponding to anticoagulant concentrations at clinical decision points in the system used by each laboratory is mandatory to be able to use them correctly in the clinic.