Automated Controlled Stirred Liquid Baths System for Calibrating Contact Thermometers in the Temperature Range from -80°C to 420°C

National Institute of Standards (NIS) is the national metrology institute in Egypt. It consists of more than twenty laboratories working continuously to realize the international standard units (SI units). Thermal metrology laboratory (ThML) is one NIS leading laboratory that has twice responsibilities, one for realizing the thermodynamic temperature and the other for introducing calibration solutions to all organizations in Egypt such as contact, non-contact thermometry, hygrometry, thermal analysis and viscosity. In this study, a newly developed automated firmware was used for controlling and regulating the temperature of stirred liquid baths and thermometry bridge system (SLBTBS) simultaneously and determined precisely the starting time of the measurements according to smart algorisms using compatible parameters of (LabVIEW®) tools. The protocol of measurements was dependent on an automatic investigation of the best stability period whereby the setting point lied in the specific temperature ranges related to hysteresis limits. Characterization tests and assessment for optimum conditions of the SLBTBS were done. The newly designed controller gave temperature stability better than 0.02K in the low-temperature range with better-linked uncertainty.

Authors

Institutions

Publication Details

Journal
Journal of Advanced Research Design
Published
2024-12-31
DOI
https://doi.org/10.37934/ard.123.1.3444
Primary Topic
Advanced Sensor Technologies Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Automated Controlled Stirred Liquid Baths System for Calibrating Contact Thermometers in the Temperature Range from -80°C to 420°C

Ahmed El-Matarawy, Eman Mohy, Doaa Mohamed
Journal of Advanced Research Design
Advanced Sensor Technologies Research
article

Automated Controlled Stirred Liquid Baths System for Calibrating Contact Thermometers in the Temperature Range from -80°C to 420°C

Ahmed El-Matarawy, Eman Mohy, Doaa Mohamed
article en

Abstract

National Institute of Standards (NIS) is the national metrology institute in Egypt. It consists of more than twenty laboratories working continuously to realize the international standard units (SI units). Thermal metrology laboratory (ThML) is one NIS leading laboratory that has twice responsibilities, one for realizing the thermodynamic temperature and the other for introducing calibration solutions to all organizations in Egypt such as contact, non-contact thermometry, hygrometry, thermal analysis and viscosity. In this study, a newly developed automated firmware was used for controlling and regulating the temperature of stirred liquid baths and thermometry bridge system (SLBTBS) simultaneously and determined precisely the starting time of the measurements according to smart algorisms using compatible parameters of (LabVIEW®) tools. The protocol of measurements was dependent on an automatic investigation of the best stability period whereby the setting point lied in the specific temperature ranges related to hysteresis limits. Characterization tests and assessment for optimum conditions of the SLBTBS were done. The newly designed controller gave temperature stability better than 0.02K in the low-temperature range with better-linked uncertainty.

Journal of Advanced Research DesignVol. 123(1)
Higher Institute of Engineering (EG), National Institute of Standards (EG)
Openalex Percentile: Top 74%
Advanced Sensor Technologies Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.