Comparative evaluation of the use of water-filled and air-charged catheters during urodynamic studies in women with stress urinary incontinence.

Introduction. Urodynamic studies play an important role in urological and gynecological clinical practice. Specialized catheters used to measure pressures inside and outside the bladder play an important role in this process. International Continence Society (ICS) standards recommend the use of water-filled catheters with external transducers for complex urodynamic studies, which ensures standardization and comparability of results between different clinicsand patients. In recent years, interest in air-charged catheters has increased as a potential alternative to traditional water-perfusion catheters catheters, supposedly reducing artifacts and improving measurement accuracy. Purpose. To conduct a comparative analysis of the data obtained during urodynamic study with the simultaneous use of air-charged and water-filled catheters in women with stress urinary incontinence. Materials and methods. A prospective comparative study of catheters for urodynamics in 41 patients with stress urinary incontinence who underwent surgical treatment at the Botkin Hospital. Air-charged and water-filled catheters were used and installed according to ICS standards. The study was performed using two urodynamic units: Duet Clinic (Mediwatch UK Ltd., Great Britain) for air-charged catheters and Dynamic Proxima (Medetron s.r.o., Czech Republic) for water-filled catheters. When comparing quantitative characteristics of two independent groups with a non-normal distribution, the Mann-Whitney U test was used, with a normal distribution – the Student test. Results. Among the main findings, air-charged catheters showed significant differences in cough pressure and Valsalva test measurements compared with water-filled catheters (p<0.05). However, differences in intravesical and abdominal pressure measurements did not reach statistical significance. Conclusion. The results of our study show that the data obtained using two types of sensors when conducting a complex urodynamic study are not completely comparable. To accurately measure bladder pressure, the use of water perfusion sensors is recommended.

Authors

Publication Details

Journal
Experimental and Сlinical Urology
Published
2025-03-25
DOI
https://doi.org/10.29188/2222-8543-2025-18-1-158-164
Primary Topic
Pelvic floor disorders treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparative evaluation of the use of water-filled and air-charged catheters during urodynamic studies in women with stress urinary incontinence.

I. O. Gritskov, George Kasyan, Pushkar' DIu, Nikita S. Kazakov
Experimental and Сlinical Urology
Pelvic floor disorders treatments
article

Comparative evaluation of the use of water-filled and air-charged catheters during urodynamic studies in women with stress urinary incontinence.

I. O. Gritskov, George Kasyan, Pushkar' DIu, Nikita S. Kazakov
article en

Abstract

Introduction. Urodynamic studies play an important role in urological and gynecological clinical practice. Specialized catheters used to measure pressures inside and outside the bladder play an important role in this process. International Continence Society (ICS) standards recommend the use of water-filled catheters with external transducers for complex urodynamic studies, which ensures standardization and comparability of results between different clinicsand patients. In recent years, interest in air-charged catheters has increased as a potential alternative to traditional water-perfusion catheters catheters, supposedly reducing artifacts and improving measurement accuracy. Purpose. To conduct a comparative analysis of the data obtained during urodynamic study with the simultaneous use of air-charged and water-filled catheters in women with stress urinary incontinence. Materials and methods. A prospective comparative study of catheters for urodynamics in 41 patients with stress urinary incontinence who underwent surgical treatment at the Botkin Hospital. Air-charged and water-filled catheters were used and installed according to ICS standards. The study was performed using two urodynamic units: Duet Clinic (Mediwatch UK Ltd., Great Britain) for air-charged catheters and Dynamic Proxima (Medetron s.r.o., Czech Republic) for water-filled catheters. When comparing quantitative characteristics of two independent groups with a non-normal distribution, the Mann-Whitney U test was used, with a normal distribution – the Student test. Results. Among the main findings, air-charged catheters showed significant differences in cough pressure and Valsalva test measurements compared with water-filled catheters (p<0.05). However, differences in intravesical and abdominal pressure measurements did not reach statistical significance. Conclusion. The results of our study show that the data obtained using two types of sensors when conducting a complex urodynamic study are not completely comparable. To accurately measure bladder pressure, the use of water perfusion sensors is recommended.

Experimental and Сlinical UrologyVol. 18(1)
Openalex Percentile: Top 84%
Pelvic floor disorders treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.