Simulation Study on the Surface Texturing Design of COC Hip Joints Based on Elastohydrodynamic Lubrication Model

Post-operative feedback from hip replacement surgeries indicates that implanted ceramic artificial hip joints may produce abnormal noises during movement. This occurrence of joint noise is highly correlated with insufficient lubrication of ceramic-on-ceramic (COC) prostheses. Studies have shown that surface texture design can improve lubrication performance. In this study, the elastohydrodynamic lubrication model was established with designing textures on the surface of the COC hip joint, using Matlab R2018b and GNU FORTRAN in Codeblocks 20.03 programming. Iterative calculations were performed to determine the average bearing capacity of the oil film and the friction coefficient. The study explored the impact of texture parameters, including the aspect ratio and density, on the lubrication and friction performance of the hip joints. The results indicate that the textured surface generally has a higher fluid film bearing capacity by 161.5~637.7% and a lower friction coefficient by 10.7~60% than the smooth surface. The average bearing capacity of the fluid film increases with an increasing texture aspect ratio, while the trend of the friction coefficient is identical to the average bearing capacity results. As the texture density increases, the average bearing capacity of the fluid film first decreases and then increases, and the trend of the friction coefficient also increases accordingly. Among the nine design groups (Sp=0.05,0.15,0.35,ε=0.075,0.1,0.15), based on the fuzzy comprehensive evaluation, the local optimal solution is Sp=0.15, ε=0.075 for lubrication and wear resistance.

Authors

Institutions

Publication Details

Journal
Lubricants
Published
2025-05-14
DOI
https://doi.org/10.3390/lubricants13050217
Citations
2
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
4.01
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Simulation Study on the Surface Texturing Design of COC Hip Joints Based on Elastohydrodynamic Lubrication Model

Zikai Hua, Leiming Gao, Zhenhua Wu, Xiuling Huang
2 citations
Lubricants
Orthopaedic implants and arthroplasty
4.01
article

Simulation Study on the Surface Texturing Design of COC Hip Joints Based on Elastohydrodynamic Lubrication Model

Zikai Hua, Leiming Gao, Zhenhua Wu, Xiuling Huang
article en
2 citations

Abstract

Post-operative feedback from hip replacement surgeries indicates that implanted ceramic artificial hip joints may produce abnormal noises during movement. This occurrence of joint noise is highly correlated with insufficient lubrication of ceramic-on-ceramic (COC) prostheses. Studies have shown that surface texture design can improve lubrication performance. In this study, the elastohydrodynamic lubrication model was established with designing textures on the surface of the COC hip joint, using Matlab R2018b and GNU FORTRAN in Codeblocks 20.03 programming. Iterative calculations were performed to determine the average bearing capacity of the oil film and the friction coefficient. The study explored the impact of texture parameters, including the aspect ratio and density, on the lubrication and friction performance of the hip joints. The results indicate that the textured surface generally has a higher fluid film bearing capacity by 161.5~637.7% and a lower friction coefficient by 10.7~60% than the smooth surface. The average bearing capacity of the fluid film increases with an increasing texture aspect ratio, while the trend of the friction coefficient is identical to the average bearing capacity results. As the texture density increases, the average bearing capacity of the fluid film first decreases and then increases, and the trend of the friction coefficient also increases accordingly. Among the nine design groups (Sp=0.05,0.15,0.35,ε=0.075,0.1,0.15), based on the fuzzy comprehensive evaluation, the local optimal solution is Sp=0.15, ε=0.075 for lubrication and wear resistance.

LubricantsVol. 13(5)
Shanghai University (CN), Nottingham Trent University (GB)
Openalex Percentile: Top 7%
Orthopaedic implants and arthroplasty
4.01
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.