In Vitro Evaluation of Accuracy of CBCT-Derived Volumes in Maxillary Defects: Effects of kVp, Device, and Software

<b>Objectives:</b> This study aimed to evaluate the accuracy of CBCT-based volumetric measurements of maxillary defects and to investigate the effects of different CBCT devices, kVp settings, and segmentation software on measurement accuracy. <b>Methods:</b> CBCT images from eight patients with maxillary defects were used to generate 3D-printed models for volumetric assessment. Two CBCT systems (Largev Smart and Planmeca Promax) were evaluated at three different kVp settings. Volume calculations were conducted using ITK-SNAP version 4.2.2 and 3D Doctor version 4.0 software, while micro-CT served as the gold standard (GS) for comparison. Statistical analysis included a three-way ANOVA to assess the effect of CBCT parameters and software on volumetric accuracy. Additionally, post-hoc Tukey HSD analysis was performed to identify specific differences between kVp groups, and Pearson correlation analysis was used to evaluate consistency with the GS. Significance level was set at <i>p</i> < 0.05. <b>Results:</b> Higher kVp settings significantly improved volumetric accuracy, with 100 kVp yielding the smallest deviations (-3.77%) from the GS. Tukey HSD analysis revealed significant differences between 60-80 kVp (<i>p</i> = 0.008), 60-100 kVp (<i>p</i> < 0.001), and 80-100 kVp (<i>p</i> = 0.041), confirming the influence of kVp on accuracy. No significant differences were observed between CBCT devices or software programs (<i>p</i> > 0.05). A strong positive correlation (r = 0.96) between CBCT-derived and micro-CT volumes confirmed CBCT's reliability for volumetric assessments (<i>p</i> < 0.001). <b>Conclusions:</b> CBCT provides accurate volumetric measurements of maxillary defects, particularly at higher kVp settings. These findings support its clinical application for preoperative planning and postoperative evaluation, offering a cost-effective alternative to micro-CT.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2025-05-14
DOI
https://doi.org/10.3390/diagnostics15101247
Primary Topic
Dental Radiography and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

In Vitro Evaluation of Accuracy of CBCT-Derived Volumes in Maxillary Defects: Effects of kVp, Device, and Software

Lorena Benítez, Leonardo Nart, Kıvanç Kamburoğlu, Wael Aboelmaaty et al.
Diagnostics
Dental Radiography and Imaging
article

In Vitro Evaluation of Accuracy of CBCT-Derived Volumes in Maxillary Defects: Effects of kVp, Device, and Software

Lorena Benítez, Leonardo Nart, Kıvanç Kamburoğlu, Wael Aboelmaaty, Sema Murat, Mohammed Awawdeh, Diego J Vázquez, Victoria Azcona
article en

Abstract

<b>Objectives:</b> This study aimed to evaluate the accuracy of CBCT-based volumetric measurements of maxillary defects and to investigate the effects of different CBCT devices, kVp settings, and segmentation software on measurement accuracy. <b>Methods:</b> CBCT images from eight patients with maxillary defects were used to generate 3D-printed models for volumetric assessment. Two CBCT systems (Largev Smart and Planmeca Promax) were evaluated at three different kVp settings. Volume calculations were conducted using ITK-SNAP version 4.2.2 and 3D Doctor version 4.0 software, while micro-CT served as the gold standard (GS) for comparison. Statistical analysis included a three-way ANOVA to assess the effect of CBCT parameters and software on volumetric accuracy. Additionally, post-hoc Tukey HSD analysis was performed to identify specific differences between kVp groups, and Pearson correlation analysis was used to evaluate consistency with the GS. Significance level was set at <i>p</i> < 0.05. <b>Results:</b> Higher kVp settings significantly improved volumetric accuracy, with 100 kVp yielding the smallest deviations (-3.77%) from the GS. Tukey HSD analysis revealed significant differences between 60-80 kVp (<i>p</i> = 0.008), 60-100 kVp (<i>p</i> < 0.001), and 80-100 kVp (<i>p</i> = 0.041), confirming the influence of kVp on accuracy. No significant differences were observed between CBCT devices or software programs (<i>p</i> > 0.05). A strong positive correlation (r = 0.96) between CBCT-derived and micro-CT volumes confirmed CBCT's reliability for volumetric assessments (<i>p</i> < 0.001). <b>Conclusions:</b> CBCT provides accurate volumetric measurements of maxillary defects, particularly at higher kVp settings. These findings support its clinical application for preoperative planning and postoperative evaluation, offering a cost-effective alternative to micro-CT.

DiagnosticsVol. 15(10)
Ankara University (TR), Mansoura University (EG), King Saud bin Abdulaziz University for Health Sciences (SA), Universidad de Buenos Aires (AR), Ahmet Yesevi University (KZ), King Abdullah International Medical Research Center (SA), National Guard Health Affairs (SA)
Openalex Percentile: Top 86%
Dental Radiography and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.