Cyclic Fatigue Resistance of “Replica-Like" and Original Reciprocating Instruments in Single and Double Curvatures
The tip sizes of the instrument tested were smaller than what is reported by the manufacturers. The cyclic fatigue resistance of the conventional 'replica-like' instrument (Reverso Silver) was significantly lower than the blue "heat-treated" comparators. Double curvature hastened fracture. (EEJ-2024-11-174).