Influence of Water Temperature in Vacuum Pumps Used for Air Removal in Steam Sterilizer

The temperature of the water used in vacuum pumps of steam sterilizers has the potential to cause failures in sterilization processes due to reduced efficiency and compromised vacuum levels. This study measured the impact of water temperature on a liquid ring vacuum pump (monobloc design) in a steam sterilizer. A sterilizer with a 566-L chamber was used for the tests. The water temperatures tested were 10°C, 20°C, 30°C, 40°C, and 50°C, and the following outcomes were analyzed: vacuum depth achieved in the drying phase (mbar), total cycle time (min), vacuum rate in the conditioning phase (mbar/s), and water consumption (L). Water temperature influenced the performance of the vacuum pump in all outcomes analyzed. Considering total cycle times, the performance with water at 10°C can be up to 50% better compared with performance with water at 50°C. The highest vacuum rates were obtained with water between 10°C and 20°C (up to 13 and 20.24 mbar/s, respectively). In cycles performed at 10°C, average water consumption was 33.3 L. In contrast, for cycles carried out at 50°C, the average water consumption increased to 94.2 L. The temperature of the water used in vacuum pumps influences the efficiency of sterilization cycles, which can result in longer cycles and higher water consumption.

Authors

Publication Details

Journal
Biomedical Instrumentation & Technology
Published
2024-10-01
DOI
https://doi.org/10.2345/0899-8205-58.4.81
Citations
2
Primary Topic
Medical Device Sterilization and Disinfection
Type
article
Field-Weighted Citation Impact
0.69
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Influence of Water Temperature in Vacuum Pumps Used for Air Removal in Steam Sterilizer

Sandoval Barbosa Rodrigues, Luís Felipe Dutra Corrêa, Rafael Queiroz de Souza
2 citations
Biomedical Instrumentation & Technology
Medical Device Sterilization and Disinfection
0.69
article

Influence of Water Temperature in Vacuum Pumps Used for Air Removal in Steam Sterilizer

Sandoval Barbosa Rodrigues, Luís Felipe Dutra Corrêa, Rafael Queiroz de Souza
article en
2 citations

Abstract

The temperature of the water used in vacuum pumps of steam sterilizers has the potential to cause failures in sterilization processes due to reduced efficiency and compromised vacuum levels. This study measured the impact of water temperature on a liquid ring vacuum pump (monobloc design) in a steam sterilizer. A sterilizer with a 566-L chamber was used for the tests. The water temperatures tested were 10°C, 20°C, 30°C, 40°C, and 50°C, and the following outcomes were analyzed: vacuum depth achieved in the drying phase (mbar), total cycle time (min), vacuum rate in the conditioning phase (mbar/s), and water consumption (L). Water temperature influenced the performance of the vacuum pump in all outcomes analyzed. Considering total cycle times, the performance with water at 10°C can be up to 50% better compared with performance with water at 50°C. The highest vacuum rates were obtained with water between 10°C and 20°C (up to 13 and 20.24 mbar/s, respectively). In cycles performed at 10°C, average water consumption was 33.3 L. In contrast, for cycles carried out at 50°C, the average water consumption increased to 94.2 L. The temperature of the water used in vacuum pumps influences the efficiency of sterilization cycles, which can result in longer cycles and higher water consumption.

Biomedical Instrumentation & TechnologyVol. 58(4)
Openalex Percentile: Top 27%
Medical Device Sterilization and Disinfection
0.69
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.