Simultaneous Sensing and Stimulation in Neuromodulation Systems: Quantifying and Mitigating Charge Accumulation Effects

This paper describes the operation of a low-noise amplifier and neurostimulation circuit when both functions are used simultaneously. The design of this circuitry along with the circuit model used to investigate system interactions is described. Expected circuit operation is explored in conjunction with the impact of charge storage at the electrode tissue interface. The test results for the system during independent use of the amplifier and during simultaneous use of the amplifier and stimulator are presented. As seen in the circuit analysis, test results confirm that the charge storage that occurs in the tissue/electrode interface used for stimulation interferes with the measurement of the neural signals of interest when the two circuits share electrodes. By optimizing the configuration used for the amplifier during high frequency stimulation, biological signals can be measured; however, the effective number of bits (ENOB) will be degraded as a function of the stimulation parameters, limiting the applications in which signal amplification can be meaningfully used when sharing an electrode with stimulation circuitry.

Authors

Institutions

Publication Details

Journal
IEEE Access
Published
2025-01-01
DOI
https://doi.org/10.1109/access.2025.3569414
Primary Topic
Neuroscience and Neural Engineering
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Simultaneous Sensing and Stimulation in Neuromodulation Systems: Quantifying and Mitigating Charge Accumulation Effects

Heather Orser, Preston Doan
IEEE Access
Neuroscience and Neural Engineering
article

Simultaneous Sensing and Stimulation in Neuromodulation Systems: Quantifying and Mitigating Charge Accumulation Effects

Heather Orser, Preston Doan
article en

Abstract

This paper describes the operation of a low-noise amplifier and neurostimulation circuit when both functions are used simultaneously. The design of this circuitry along with the circuit model used to investigate system interactions is described. Expected circuit operation is explored in conjunction with the impact of charge storage at the electrode tissue interface. The test results for the system during independent use of the amplifier and during simultaneous use of the amplifier and stimulator are presented. As seen in the circuit analysis, test results confirm that the charge storage that occurs in the tissue/electrode interface used for stimulation interferes with the measurement of the neural signals of interest when the two circuits share electrodes. By optimizing the configuration used for the amplifier during high frequency stimulation, biological signals can be measured; however, the effective number of bits (ENOB) will be degraded as a function of the stimulation parameters, limiting the applications in which signal amplification can be meaningfully used when sharing an electrode with stimulation circuitry.

IEEE AccessVol. 13
University of St. Thomas - Minnesota (US), The Milton H. Erickson Foundation (US)
Openalex Percentile: Top 88%
Neuroscience and Neural Engineering
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.