Using a microfluidic paper-based analytical device and solid-phase extraction to determine phosphate concentration

Phosphate is an essential nutrient, but in high concentrations it contributes to water pollution. Traditional methods for phosphate measurement, such as absorption spectrophotometry and ion chromatography, require expensive equipment and skilled operators. This study introduces a microfluidic paper-based analytical device (μPAD) that is designed to accomplish field-based, low-concentration phosphate measurements. This μPAD utilizes colorimetric detection based on the molybdenum blue method. Herein, we describe how the conditions were optimized in terms of design and sensitivity by adjusting reagent concentrations, paper thickness, and the time frames for sample introduction, and reaction. The operation consists of simply dipping the μPAD into a sample, capturing images in a home-made photo studio box, and processing the images with ImageJ software to measure RGB intensity. An additional preconcentration step involves solid-phase extraction with an anion exchange resin that achieves a 10-fold enrichment, which enables detection that ranges from 0.05 to 1 mg L<sup>-1</sup> with a detection limit of 0.089 mg L<sup>-1</sup> and a quantification limit of 0.269 mg L<sup>-1</sup>. The replicated measurements showed good reproducibility both intraday and interday (five different days) as 4.7 % and 3.0 % of relative standard deviations, respectively. After storage in a refrigerator for as long as 26 days, this μPAD delivered stable and accurate results for real-world samples of natural water, soil, and toothpaste. The results produced using this system correlate well with those produced via spectrophotometry. This μPAD-based method is a cost-effective, portable, rapid, and simple approach that allows relatively unskilled operators to monitor phosphate concentrations in field applications.

Authors

Institutions

Publication Details

Journal
Talanta
Published
2025-05-10
DOI
https://doi.org/10.1016/j.talanta.2025.128303
Citations
4
Primary Topic
Biosensors and Analytical Detection
Type
article
Field-Weighted Citation Impact
2.70
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Using a microfluidic paper-based analytical device and solid-phase extraction to determine phosphate concentration

Kaewta Danchana, Haruka Namba, Takashi Kaneta
4 citations
Talanta
Biosensors and Analytical Detection
2.70
article

Using a microfluidic paper-based analytical device and solid-phase extraction to determine phosphate concentration

Kaewta Danchana, Haruka Namba, Takashi Kaneta
article en
4 citations

Abstract

Phosphate is an essential nutrient, but in high concentrations it contributes to water pollution. Traditional methods for phosphate measurement, such as absorption spectrophotometry and ion chromatography, require expensive equipment and skilled operators. This study introduces a microfluidic paper-based analytical device (μPAD) that is designed to accomplish field-based, low-concentration phosphate measurements. This μPAD utilizes colorimetric detection based on the molybdenum blue method. Herein, we describe how the conditions were optimized in terms of design and sensitivity by adjusting reagent concentrations, paper thickness, and the time frames for sample introduction, and reaction. The operation consists of simply dipping the μPAD into a sample, capturing images in a home-made photo studio box, and processing the images with ImageJ software to measure RGB intensity. An additional preconcentration step involves solid-phase extraction with an anion exchange resin that achieves a 10-fold enrichment, which enables detection that ranges from 0.05 to 1 mg L<sup>-1</sup> with a detection limit of 0.089 mg L<sup>-1</sup> and a quantification limit of 0.269 mg L<sup>-1</sup>. The replicated measurements showed good reproducibility both intraday and interday (five different days) as 4.7 % and 3.0 % of relative standard deviations, respectively. After storage in a refrigerator for as long as 26 days, this μPAD delivered stable and accurate results for real-world samples of natural water, soil, and toothpaste. The results produced using this system correlate well with those produced via spectrophotometry. This μPAD-based method is a cost-effective, portable, rapid, and simple approach that allows relatively unskilled operators to monitor phosphate concentrations in field applications.

TalantaVol. 295
Okayama University of Science (JP), Okayama University (JP)
Openalex Percentile: Top 10%
Biosensors and Analytical Detection
2.70
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Using a microfluidic paper-based analytical device and solid-phase extraction to determine phosphate concentration — Kaewta Danchana, Haruka Namba, et al. · Talanta (2025) | TGRS Research Map | TGRS