Quantitative capabilities of commercial CZT SPECT-CT cameras with 99mTc
Without scatter correction, the CZT SPECT systems showed contrast performance similar to the NaI camera. With scatter correction, this held true for cold objects but the contrast of hot objects was not significantly improved or was degraded depending on the system considered and the object size. Quantification errors of less than 5% were achievable. It is hoped that on-going developments at both manufacturers will improve the scatter correction accuracy.