Is a 3-Minute Knee MRI Protocol Sufficient for Daily Clinical Practice? A SuperResolution Reconstruction Approach Using AI and Compressed Sensing

<b>Objectives:</b> The purpose of this study was to assess whether a 3-min 2D knee protocol can meet the needs for clinical application if using a SuperResolution reconstruction approach. <b>Methods:</b> In this prospective study, a total of 20 volunteers underwent imaging of the knee using a 3T MRI scanner (Philips Ingenia Elition X 3.0T, Philips). The imaging protocol, consisting of a fat-saturated 2D proton density sequence in coronal, sagittal, and transverse orientations, as well as a sagittal T1-weighted sequence, was acquired with standard and ultra-low resolution. The standard sequences were reconstructed using an AI-assisted Compressed SENSE method (SmartSpeed). The ultra-low-resolution sequences have been reconstructed using a vendor-provided prototype. Four experienced readers (two radiologists and two orthopedic surgeons) evaluated the sequences for image quality, anatomical structures, and incidental pathologies. The consensus evaluation of two different experienced radiologists specialized in musculoskeletal imaging served as the gold standard. <b>Results:</b> The acquisition time for the entire protocol was 11:01 min for standard resolution and 03:36 min for ultra-low resolution. In the overall assessment, CS-SuperRes-reconstructed sequences showed slightly improved accuracy and increased specificity compared to the standard CS-AI method (0.87 vs. 0.86 and 0.9 vs. 0.87, respectively), while the standard method exhibited a higher sensitivity (0.73 vs. 0.57). Overall, 24 out of 40 pathologies were detected in the ultra-low-resolution images compared to 26 in the standard images. <b>Conclusions:</b> The CS-SuperRes method enables a 2D knee protocol to be completed in 3 min, with improved accuracy compared to the clinical standard.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2025-05-09
DOI
https://doi.org/10.3390/diagnostics15101206
Citations
4
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
5.44
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Is a 3-Minute Knee MRI Protocol Sufficient for Daily Clinical Practice? A SuperResolution Reconstruction Approach Using AI and Compressed Sensing

Philip Rauen, Lajos Basten, Robert Hahnfeldt, Andra-Iza Iuga et al.
4 citations
Diagnostics
Advanced MRI Techniques and Applications
5.44
article

Is a 3-Minute Knee MRI Protocol Sufficient for Daily Clinical Practice? A SuperResolution Reconstruction Approach Using AI and Compressed Sensing

Philip Rauen, Lajos Basten, Robert Hahnfeldt, Andra-Iza Iuga, Johannes Oppermann, Robert Terzis, Jan Paul Janßen, Kristina Sonnabend, Jonathan Kottlors, Grischa Bratke, Thomas Dratsch, Christoph Katemann, Stephan Skornitzke, David Grevenstein, David Maintz, Nour El-Hoda Abou Zeid
article en
4 citations

Abstract

<b>Objectives:</b> The purpose of this study was to assess whether a 3-min 2D knee protocol can meet the needs for clinical application if using a SuperResolution reconstruction approach. <b>Methods:</b> In this prospective study, a total of 20 volunteers underwent imaging of the knee using a 3T MRI scanner (Philips Ingenia Elition X 3.0T, Philips). The imaging protocol, consisting of a fat-saturated 2D proton density sequence in coronal, sagittal, and transverse orientations, as well as a sagittal T1-weighted sequence, was acquired with standard and ultra-low resolution. The standard sequences were reconstructed using an AI-assisted Compressed SENSE method (SmartSpeed). The ultra-low-resolution sequences have been reconstructed using a vendor-provided prototype. Four experienced readers (two radiologists and two orthopedic surgeons) evaluated the sequences for image quality, anatomical structures, and incidental pathologies. The consensus evaluation of two different experienced radiologists specialized in musculoskeletal imaging served as the gold standard. <b>Results:</b> The acquisition time for the entire protocol was 11:01 min for standard resolution and 03:36 min for ultra-low resolution. In the overall assessment, CS-SuperRes-reconstructed sequences showed slightly improved accuracy and increased specificity compared to the standard CS-AI method (0.87 vs. 0.86 and 0.9 vs. 0.87, respectively), while the standard method exhibited a higher sensitivity (0.73 vs. 0.57). Overall, 24 out of 40 pathologies were detected in the ultra-low-resolution images compared to 26 in the standard images. <b>Conclusions:</b> The CS-SuperRes method enables a 2D knee protocol to be completed in 3 min, with improved accuracy compared to the clinical standard.

DiagnosticsVol. 15(10)
University of Cologne (DE), University Hospital Cologne (DE), Orthopädische Praxis (DE), Philips (Germany) (DE)
Openalex Percentile: Top 4%
Advanced MRI Techniques and Applications
5.44
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.