Comparison of two different autotransfusion devices: An ex-vivo study

IntroductionDuring surgery, blood is recovered from the surgical field and the autotransfusion device separates and washes the red blood cells (RBC's) after which these can be retransfused to the patient. Autotransfusion devices differ strongly in separation method and washing settings, which may lead to different RBC recovery rates and removal rates of contaminants. The objective of this study was to compare the autoLog IQ<sup>TM</sup> (Medtronic) and the Xtra<sup>®</sup> (LivaNova) in terms of RBC recovery, quality of the processed blood and processing time.MethodsHuman blood was heparinised and processed with both autotransfusion devices according to their standard protocols. Blood samples were taken from the collection reservoir and from the transfusion bag and analyzed for cell count, heparin removal and cell injury.ResultsThe RBC recovery rates for both devices was 93%. Heparin was almost completely removed in both devices with >99.9%. The autoLog IQ<sup>TM</sup> showed significantly better removal of platelets (autoLog IQ vs Xtra; 86.1 ± 2.7 and 78.6 ± 4.4%, <i>p</i> = 0.01) and of leukocytes (autoLog IQ vs Xtra; 39.6 ± 2.7 and 13.4 ± 5.7%, <i>p</i> < 0.001). No other significant differences were observed in removal rates. The volume of RBC concentrated per minute was faster for the Xtra<sup>®</sup> (autoLog IQ vs Xtra, 21 ± 3 and 27 ± 2 mL RBC/min, <i>p</i> = 0.007).ConclusionsIn this study both the autoLog IQ<sup>TM</sup> and the Xtra<sup>®</sup> showed similar RBC recovery rates of 93% and almost all heparin was eliminated. The washing quality of the autoLog IQ<sup>TM</sup> device appears to be better, with better removal of platelets and leukocytes. Although both the autoLog IQ<sup>TM</sup> and the Xtra<sup>®</sup> devices use very different separation techniques and washing protocols, the difference in processing speed and various markers for cell damage in the end product seems trivial.

Authors

Institutions

Publication Details

Journal
Perfusion
Published
2025-05-07
DOI
https://doi.org/10.1177/02676591251340938
Primary Topic
Blood transfusion and management
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of two different autotransfusion devices: An ex-vivo study

Marco C. Stehouwer, Sander van Deventer, Wim van Oeveren
Perfusion
Blood transfusion and management
article

Comparison of two different autotransfusion devices: An ex-vivo study

Marco C. Stehouwer, Sander van Deventer, Wim van Oeveren
article en

Abstract

IntroductionDuring surgery, blood is recovered from the surgical field and the autotransfusion device separates and washes the red blood cells (RBC's) after which these can be retransfused to the patient. Autotransfusion devices differ strongly in separation method and washing settings, which may lead to different RBC recovery rates and removal rates of contaminants. The objective of this study was to compare the autoLog IQ<sup>TM</sup> (Medtronic) and the Xtra<sup>®</sup> (LivaNova) in terms of RBC recovery, quality of the processed blood and processing time.MethodsHuman blood was heparinised and processed with both autotransfusion devices according to their standard protocols. Blood samples were taken from the collection reservoir and from the transfusion bag and analyzed for cell count, heparin removal and cell injury.ResultsThe RBC recovery rates for both devices was 93%. Heparin was almost completely removed in both devices with >99.9%. The autoLog IQ<sup>TM</sup> showed significantly better removal of platelets (autoLog IQ vs Xtra; 86.1 ± 2.7 and 78.6 ± 4.4%, <i>p</i> = 0.01) and of leukocytes (autoLog IQ vs Xtra; 39.6 ± 2.7 and 13.4 ± 5.7%, <i>p</i> < 0.001). No other significant differences were observed in removal rates. The volume of RBC concentrated per minute was faster for the Xtra<sup>®</sup> (autoLog IQ vs Xtra, 21 ± 3 and 27 ± 2 mL RBC/min, <i>p</i> = 0.007).ConclusionsIn this study both the autoLog IQ<sup>TM</sup> and the Xtra<sup>®</sup> showed similar RBC recovery rates of 93% and almost all heparin was eliminated. The washing quality of the autoLog IQ<sup>TM</sup> device appears to be better, with better removal of platelets and leukocytes. Although both the autoLog IQ<sup>TM</sup> and the Xtra<sup>®</sup> devices use very different separation techniques and washing protocols, the difference in processing speed and various markers for cell damage in the end product seems trivial.

PerfusionVol. 41(1)
Hashemite University (JO), St. Antonius Ziekenhuis (NL)
Medtronic Europe
Openalex Percentile: Top 91%
Blood transfusion and management
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.