Metrological-Characteristics-Based Calibration of Optical Areal Surface Measuring Instruments and Evaluation of Measurement Uncertainty for Surface Texture Measurements
ISO 25178 part 600:2019 and part 700:2022 introduce a calibration framework based on seven metrological characteristics (MCs) for calibrating optical areal surface measuring instruments. Among these, topography fidelity is a newly defined metrological characteristic that remains a critical yet unresolved challenge in instrument calibration. This paper proposes strategies to address topography fidelity, including a key criterion for selecting suitable instrument setups by comparing slope measurement capability with local surface slopes, as well as methods for investigating the field-of-view homogeneity and directional performance difference along the x- and y-axes. Furthermore, the uncertainty contribution of topography fidelity in surface topography measurements is analysed. The paper also determines the uncertainty associated with the remaining six MCs. Based on the proposed MC-based calibration approach and the corresponding uncertainty contributions, an overall measurement uncertainty model for Sa and Sq parameters is presented. Finally, uncertainty evaluations for Sa and Sq are demonstrated on a challenging surface, where topography fidelity plays a significant role in the measurement uncertainty evaluation.
Authors
- Uwe Brand (ORCID: https://orcid.org/0000-0002-4293-4982)
- André Felgner (ORCID: https://orcid.org/0000-0002-7851-2220)
- Sai Gao (ORCID: https://orcid.org/0009-0004-1465-9524)
Institutions
- Physikalisch-Technische Bundesanstalt (DE)
Publication Details
- Journal
- Instruments
- Published
- 2025-05-08
- DOI
- https://doi.org/10.3390/instruments9020011
- Citations
- 3
- Primary Topic
- Surface Roughness and Optical Measurements
- Type
- article
- Field-Weighted Citation Impact
- 4.00
Funders
- European Commission
- European Metrology Programme for Innovation and Research