Simulation and experimental study on processing behavior of coronary artery calcified tissue removal

Coronary artery atherosclerosis is a prevalent cardiovascular disease and a leading cause of major adverse cardiovascular events (MACE). Rotational atherectomy (RA) is an effective interventional technique for treating severe calcified stenosis. However, excessive forces, heat, and debris are prone to lead to serious surgical complications, such as slow flow/no-reflow and blood clots. To mitigate excessive force and heat generation during RA, a novel high-performance cutting tool was designed and fabricated for coronary artery calcified tissue removal. An RA simulation model was developed to simulate the procedure. The results showed that the forces, temperatures, and debris size remained within predefined safety thresholds. Using the 1.5 mm tool as an illustration, the peak cutting force was 1.062 N, and the peak temperature rise reached 1.170 °C. Debris distribution exhibited a normal pattern, with 90% of particles measuring below 14 μm. The experimental results closely matched the simulation values, showcasing errors under 10% and affirming the simulation model's precision. This research provides theoretical support for the study of mechanisms and contributes to optimizing the effectiveness of RA.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2025-05-08
DOI
https://doi.org/10.1038/s41598-025-01236-3
Primary Topic
Coronary Interventions and Diagnostics
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Simulation and experimental study on processing behavior of coronary artery calcified tissue removal

Zhaoju Zhu, Fan Wu, Mingcheng Fang, Bingwei He et al.
Scientific Reports
Coronary Interventions and Diagnostics
article

Simulation and experimental study on processing behavior of coronary artery calcified tissue removal

Zhaoju Zhu, Fan Wu, Mingcheng Fang, Bingwei He, Chuhang Gao, Jialiang Zhu, Ziyu Cui
article en

Abstract

Coronary artery atherosclerosis is a prevalent cardiovascular disease and a leading cause of major adverse cardiovascular events (MACE). Rotational atherectomy (RA) is an effective interventional technique for treating severe calcified stenosis. However, excessive forces, heat, and debris are prone to lead to serious surgical complications, such as slow flow/no-reflow and blood clots. To mitigate excessive force and heat generation during RA, a novel high-performance cutting tool was designed and fabricated for coronary artery calcified tissue removal. An RA simulation model was developed to simulate the procedure. The results showed that the forces, temperatures, and debris size remained within predefined safety thresholds. Using the 1.5 mm tool as an illustration, the peak cutting force was 1.062 N, and the peak temperature rise reached 1.170 °C. Debris distribution exhibited a normal pattern, with 90% of particles measuring below 14 μm. The experimental results closely matched the simulation values, showcasing errors under 10% and affirming the simulation model's precision. This research provides theoretical support for the study of mechanisms and contributes to optimizing the effectiveness of RA.

Scientific ReportsVol. 15(1)
Fujian Medical University (CN), Fuzhou University (CN)
National Natural Science Foundation of China, Fujian Medical University, Fujian Provincial Health Technology Project
Openalex Percentile: Top 83%
Coronary Interventions and Diagnostics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.