Design, Test, and Shimming of a 1.5-T Head-Only MRI Superconducting Magnet

As the core component of magnetic resonance imaging (MRI) system, superconducting magnet provides a high-intensity, stable and homogeneous magnetic field background for the system. The strength, stability, and homogeneity of the magnetic field directly affect the imaging quality of MRI. This paper presents the design, test and shimming of a 1.5 T Head-Only MRI superconducting magnet developed by the Wuhan National High Magnetic Field Center (WHMFC). The magnet consists of 7 primary coils and 2 shielding coils and is operated at an operating current of 510 A using liquid helium (LHe) bath cooling. The test has shown that the magnet reached the target field of 1.5 T directly after the first ramp-up without any quench. Under the magnetic field of 1.5 T, the magnetic field decay rate of the magnet is less than 0.02 ppm/hr. After 3 rounds of passive shimming, the magnet achieves a magnetic field inhomogeneity of 2.89 ppm over an imaging sphere within a diameter of 26 cm. The successful development of this magnet provides practical engineering validation for the lightweight technology of head-only MRI superconducting magnets.

Authors

Institutions

Publication Details

Journal
IEEE Transactions on Applied Superconductivity
Published
2025-05-08
DOI
https://doi.org/10.1109/tasc.2025.3568142
Citations
1
Primary Topic
Superconducting Materials and Applications
Type
article
Field-Weighted Citation Impact
0.67

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Design, Test, and Shimming of a 1.5-T Head-Only MRI Superconducting Magnet

Jianglan Li, Yunxing Song, Liang Li, Xia Jiang et al.
1 citations
IEEE Transactions on Applied Superconductivity
Superconducting Materials and Applications
0.67
article

Design, Test, and Shimming of a 1.5-T Head-Only MRI Superconducting Magnet

Jianglan Li, Yunxing Song, Liang Li, Xia Jiang, Zhiwen Cheng, Qiuliang Wang
article en
1 citations

Abstract

As the core component of magnetic resonance imaging (MRI) system, superconducting magnet provides a high-intensity, stable and homogeneous magnetic field background for the system. The strength, stability, and homogeneity of the magnetic field directly affect the imaging quality of MRI. This paper presents the design, test and shimming of a 1.5 T Head-Only MRI superconducting magnet developed by the Wuhan National High Magnetic Field Center (WHMFC). The magnet consists of 7 primary coils and 2 shielding coils and is operated at an operating current of 510 A using liquid helium (LHe) bath cooling. The test has shown that the magnet reached the target field of 1.5 T directly after the first ramp-up without any quench. Under the magnetic field of 1.5 T, the magnetic field decay rate of the magnet is less than 0.02 ppm/hr. After 3 rounds of passive shimming, the magnet achieves a magnetic field inhomogeneity of 2.89 ppm over an imaging sphere within a diameter of 26 cm. The successful development of this magnet provides practical engineering validation for the lightweight technology of head-only MRI superconducting magnets.

IEEE Transactions on Applied SuperconductivityVol. 35(6)
Chinese Academy of Sciences (CN), Institute of Electrical Engineering (CN), Wuhan National Laboratory for Optoelectronics (CN), Huazhong University of Science and Technology (CN)
National Natural Science Foundation of China
Affordable and clean energy
Openalex Percentile: Top 32%
Superconducting Materials and Applications
0.67
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.