Dosimetric commissioning of small animal FLASH radiation research platform

<i>Objective.</i>The FLASH-SARRP, a new small animal radiation research platform has been designed to support conventional, high and ultrahigh dose-rate kV x-rays for preclinical research. This self-shielded system features two high-capacity x-ray sources with rotating-anode technology. This study characterizes the dosimetric and mechanical performances of the system for preclinical FLASH radiation research.<i>Approach.</i>Mechanical alignment of two x-ray tubes was performed using a custom-designed jig by aligning the outlet ports of the tube housings. Alignment of mechanical and radiation centers was evaluated by scanning a highly-collimated slit across the focal-spot. The linearity of the x-ray tube voltage, current and exposure-time was evaluated using silicon diode and ion-chamber detectors. Dosimetric characteristics of beam e.g. output linearity, depth dose-rate and profiles were measured using calibrated radiochromic films, thermoluminescence, and ion-chamber detectors in kV solid-water phantom or air, with and without external energy filtration. Dose-rate uniformity, flatness, symmetry, beam width, and penumbra were assessed for single and parallel-opposed x-ray beams across various field sizes.<i>Results.</i>The x-ray sources were aligned at 0.3 mm accuracy. The radiation beam center was within 1.0 mm of mechanical center. Beam output was highly linear with wide ranges of tube current (5-630 mA) and exposure-time (5-6300 ms), supporting accurate dose-rate and dose adjustments. The FLASH-SARRP supports a wide range of dose-rates from <1 Gy s<sup>-1</sup>to 100 Gy s<sup>-1</sup>, depending on field size. The uniformity of the depth and crossbeam dose-rates is ±3.6 Gy s<sup>-1</sup>and ±1.5 Gy s<sup>-1</sup>between 5-15 mm phantom depth without and with external filter, respectively.<i>Significance.</i>The FLASH-SARRP provides desirable dosimetric performance for small animal irradiation, supporting both conventional and FLASH dose-rate across field sizes from 5 mm-diameter circular to 20 mm-square apertures. This platform enables comparative studies between FLASH and conventional dose-rates in small animal (e.g. mouse) models.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-05-08
DOI
https://doi.org/10.1088/1361-6560/add641
Citations
4
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
6.07
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Dosimetric commissioning of small animal FLASH radiation research platform

Mohammad Ali Tajik‐Mansoury, Daniel Sforza, Iulian Iordachita, Mohammad Rezaee et al.
4 citations
Physics in Medicine and Biology
Radiation Therapy and Dosimetry
6.07
article

Dosimetric commissioning of small animal FLASH radiation research platform

Mohammad Ali Tajik‐Mansoury, Daniel Sforza, Iulian Iordachita, Mohammad Rezaee, John W. Wong
article en
4 citations

Abstract

<i>Objective.</i>The FLASH-SARRP, a new small animal radiation research platform has been designed to support conventional, high and ultrahigh dose-rate kV x-rays for preclinical research. This self-shielded system features two high-capacity x-ray sources with rotating-anode technology. This study characterizes the dosimetric and mechanical performances of the system for preclinical FLASH radiation research.<i>Approach.</i>Mechanical alignment of two x-ray tubes was performed using a custom-designed jig by aligning the outlet ports of the tube housings. Alignment of mechanical and radiation centers was evaluated by scanning a highly-collimated slit across the focal-spot. The linearity of the x-ray tube voltage, current and exposure-time was evaluated using silicon diode and ion-chamber detectors. Dosimetric characteristics of beam e.g. output linearity, depth dose-rate and profiles were measured using calibrated radiochromic films, thermoluminescence, and ion-chamber detectors in kV solid-water phantom or air, with and without external energy filtration. Dose-rate uniformity, flatness, symmetry, beam width, and penumbra were assessed for single and parallel-opposed x-ray beams across various field sizes.<i>Results.</i>The x-ray sources were aligned at 0.3 mm accuracy. The radiation beam center was within 1.0 mm of mechanical center. Beam output was highly linear with wide ranges of tube current (5-630 mA) and exposure-time (5-6300 ms), supporting accurate dose-rate and dose adjustments. The FLASH-SARRP supports a wide range of dose-rates from <1 Gy s<sup>-1</sup>to 100 Gy s<sup>-1</sup>, depending on field size. The uniformity of the depth and crossbeam dose-rates is ±3.6 Gy s<sup>-1</sup>and ±1.5 Gy s<sup>-1</sup>between 5-15 mm phantom depth without and with external filter, respectively.<i>Significance.</i>The FLASH-SARRP provides desirable dosimetric performance for small animal irradiation, supporting both conventional and FLASH dose-rate across field sizes from 5 mm-diameter circular to 20 mm-square apertures. This platform enables comparative studies between FLASH and conventional dose-rates in small animal (e.g. mouse) models.

Physics in Medicine and BiologyVol. 70(11)
Johns Hopkins University (US), Johns Hopkins Medicine (US), University of Baltimore (US)
Affordable and clean energy
Openalex Percentile: Top 4%
Radiation Therapy and Dosimetry
6.07
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.