The Reliability and Validity of an Isometric Knee Strength Measurement Device in Older Adult Individuals
This study aims to evaluate the reliability and validity of the Leg Strength Analyzer (IB-LS) in assessing isometric knee flexion and extension strength in elderly adults, compare its performance with that of the CSMI dynamometer, and examine its agreement with isokinetic knee strength measurements. A total of 21 elderly participants (mean age: 65.10 ± 4.56 years) were recruited. Participants underwent knee flexion and extension strength assessments using both the IB-LS and CSMI devices, with isokinetic strength at 60°/s measured in a follow-up session at least one day later. The IB-LS demonstrated high test-retest reliability in elderly adults (ICC = 0.856-0.987). The validity analysis comparing IB-LS isometric peak torque with CSMI isometric peak torque showed moderate to high validity (ICC = 0.826-0.946). Furthermore, IB-LS isometric peak torque and CSMI isokinetic 60°/s peak torque demonstrated high agreement (ICC = 0.775-0.881), demonstrating its strong association with isokinetic strength assessments. Bland-Altman analysis revealed that mean differences between IB-LS and CSMI isometric peak torque values ranged from 13.2 to 93.5 Nm, with limits of agreement (LoA) spanning from -55.8 to 192.5 Nm. When comparing IB-LS isometric peak torque with CSMI isokinetic 60°/s peak torque, mean differences ranged from 31.0 to 56.0 Nm, with LoA from -28.5 to 138.9 Nm. The IB-LS is a reliable and valid tool for evaluating isometric knee strength in elderly adults. Its strong agreement with the CSMI dynamometer and close correlation with isokinetic strength measurements indicate that IB-LS can be a feasible alternative for assessing knee strength in clinical and research settings focused on elderly populations.
Authors
- Dae-Sung Park (ORCID: https://orcid.org/0000-0003-4258-0878)
- Byoung-Kwon Lee
- Jeong-Bae Ko
- Su-Ha Lee (ORCID: https://orcid.org/0000-0001-8289-5626)
- Jae Soo Hong (ORCID: https://orcid.org/0000-0001-8564-8647)
- Ming‐Shaung Ju (ORCID: https://orcid.org/0000-0002-1297-5150)
Institutions
- Konyang University (KR)
- Sahmyook University (KR)
- Korea Institute of Industrial Technology (KR)
Publication Details
- Journal
- Sensors
- Published
- 2025-05-08
- DOI
- https://doi.org/10.3390/s25102981
- Citations
- 5
- Primary Topic
- Muscle activation and electromyography studies
- Type
- article
- Field-Weighted Citation Impact
- 3.37
Funders
- Korea Institute of Industrial Technology