The Reliability and Validity of an Isometric Knee Strength Measurement Device in Older Adult Individuals

This study aims to evaluate the reliability and validity of the Leg Strength Analyzer (IB-LS) in assessing isometric knee flexion and extension strength in elderly adults, compare its performance with that of the CSMI dynamometer, and examine its agreement with isokinetic knee strength measurements. A total of 21 elderly participants (mean age: 65.10 ± 4.56 years) were recruited. Participants underwent knee flexion and extension strength assessments using both the IB-LS and CSMI devices, with isokinetic strength at 60°/s measured in a follow-up session at least one day later. The IB-LS demonstrated high test-retest reliability in elderly adults (ICC = 0.856-0.987). The validity analysis comparing IB-LS isometric peak torque with CSMI isometric peak torque showed moderate to high validity (ICC = 0.826-0.946). Furthermore, IB-LS isometric peak torque and CSMI isokinetic 60°/s peak torque demonstrated high agreement (ICC = 0.775-0.881), demonstrating its strong association with isokinetic strength assessments. Bland-Altman analysis revealed that mean differences between IB-LS and CSMI isometric peak torque values ranged from 13.2 to 93.5 Nm, with limits of agreement (LoA) spanning from -55.8 to 192.5 Nm. When comparing IB-LS isometric peak torque with CSMI isokinetic 60°/s peak torque, mean differences ranged from 31.0 to 56.0 Nm, with LoA from -28.5 to 138.9 Nm. The IB-LS is a reliable and valid tool for evaluating isometric knee strength in elderly adults. Its strong agreement with the CSMI dynamometer and close correlation with isokinetic strength measurements indicate that IB-LS can be a feasible alternative for assessing knee strength in clinical and research settings focused on elderly populations.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2025-05-08
DOI
https://doi.org/10.3390/s25102981
Citations
5
Primary Topic
Muscle activation and electromyography studies
Type
article
Field-Weighted Citation Impact
3.37

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The Reliability and Validity of an Isometric Knee Strength Measurement Device in Older Adult Individuals

Dae-Sung Park, Byoung-Kwon Lee, Jeong-Bae Ko, Su-Ha Lee et al.
5 citations
Sensors
Muscle activation and electromyography studies
3.37
article

The Reliability and Validity of an Isometric Knee Strength Measurement Device in Older Adult Individuals

Dae-Sung Park, Byoung-Kwon Lee, Jeong-Bae Ko, Su-Ha Lee, Jae Soo Hong, Ming‐Shaung Ju
article en
5 citations

Abstract

This study aims to evaluate the reliability and validity of the Leg Strength Analyzer (IB-LS) in assessing isometric knee flexion and extension strength in elderly adults, compare its performance with that of the CSMI dynamometer, and examine its agreement with isokinetic knee strength measurements. A total of 21 elderly participants (mean age: 65.10 ± 4.56 years) were recruited. Participants underwent knee flexion and extension strength assessments using both the IB-LS and CSMI devices, with isokinetic strength at 60°/s measured in a follow-up session at least one day later. The IB-LS demonstrated high test-retest reliability in elderly adults (ICC = 0.856-0.987). The validity analysis comparing IB-LS isometric peak torque with CSMI isometric peak torque showed moderate to high validity (ICC = 0.826-0.946). Furthermore, IB-LS isometric peak torque and CSMI isokinetic 60°/s peak torque demonstrated high agreement (ICC = 0.775-0.881), demonstrating its strong association with isokinetic strength assessments. Bland-Altman analysis revealed that mean differences between IB-LS and CSMI isometric peak torque values ranged from 13.2 to 93.5 Nm, with limits of agreement (LoA) spanning from -55.8 to 192.5 Nm. When comparing IB-LS isometric peak torque with CSMI isokinetic 60°/s peak torque, mean differences ranged from 31.0 to 56.0 Nm, with LoA from -28.5 to 138.9 Nm. The IB-LS is a reliable and valid tool for evaluating isometric knee strength in elderly adults. Its strong agreement with the CSMI dynamometer and close correlation with isokinetic strength measurements indicate that IB-LS can be a feasible alternative for assessing knee strength in clinical and research settings focused on elderly populations.

SensorsVol. 25(10)
Konyang University (KR), Sahmyook University (KR), Korea Institute of Industrial Technology (KR)
Korea Institute of Industrial Technology
Openalex Percentile: Top 8%
Muscle activation and electromyography studies
3.37
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.