Evaluation of the Current State of Thyroid Hormone Testing in Human Serum—Results of the Free Thyroxine and Thyrotropin Interlaboratory Comparison Study

<b><i>Background:</i></b> Performance of thyroid function assays can vary significantly. To address this issue, the Centers for Disease Control and Prevention (CDC) Clinical Standardization Programs conducted an interlaboratory comparison of free thyroxine (fT4) immunoassays (IAs) and laboratory-developed tests (LDTs). This assessment aimed to determine the current performance characteristics of these assays as a first step toward measurement standardization. Thyrotropin (TSH) IAs were also evaluated. <b><i>Methods:</i></b> Assays measured 41 blinded individual-donor sera, including a sample from a pregnant woman (for fT4 analysis only) and three serum pools, with 11.3-32.1 pmol/L (0.881-2.49 ng/dL) fT4 and 0.337-21.6 mIU/L TSH in duplicate over 2 days. Passing-Bablok regression analysis performed pre-recalibration compared assays performance to the CDC fT4 reference measurement procedure (RMP) or TSH all-lab mean (ALM). Additionally, the impact of linear regression-based recalibration of assays to the CDC fT4 RMP or TSH ALM was estimated. Inter-assay agreement of sample classification according to the assay-specific reference interval (RI) was assessed pre- and post-recalibration. <b><i>Results:</i></b> A total of 21 fT4 and 17 TSH assays participated. Pre-recalibration, median biases of TSH measurements to the ALM were -1.2% [confidence interval or CI -1.8% to -0.4%], and good classification agreement among TSH assays was observed. fT4 assays all showed a negative median bias to the RMP, with higher bias among IAs (median: -20.3%, CI [-21.5% to -19.4%]) than LDTs (median: -4.5%, [CI -6.1% to -3.2%]). Of the individual-donor sera, only 21 out of 40 samples were classified uniformly by all fT4 assays, indicating poor inter-assay agreement. Post-recalibration, agreement improved to 33 out of 40 individual-donor sera correctly classified by all tested IAs and LDTs. Similar improvement in post-recalibration median percent bias was observed for fT4 IAs (median: -0.2, [CI -1.2% to 0.6%]) and LDTs (median: -0.3%, [CI -2.5% to 1.4%]). <b><i>Conclusions:</i></b> The comparison among fT4 assays emphasizes the need for measurement standardization to improve accuracy and comparability. This and previous studies demonstrate the possibility to develop common fT4 RIs via standardization, enabling the use of evidence-based clinical guidelines universally in patient care. Recalibration can effectively address high variability in fT4 assays, ensuring consistent diagnostic classification.

Authors

Institutions

Publication Details

Journal
Thyroid
Published
2025-05-01
DOI
https://doi.org/10.1089/thy.2024.0728
Citations
8
Primary Topic
Thyroid Disorders and Treatments
Type
article
Field-Weighted Citation Impact
10.36
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of the Current State of Thyroid Hormone Testing in Human Serum—Results of the Free Thyroxine and Thyrotropin Interlaboratory Comparison Study

Uliana Danilenko, Ashley Ribera, Alicia N. Lyle, Tatiana Buchannan et al.
8 citations
Thyroid
Thyroid Disorders and Treatments
10.36
article

Evaluation of the Current State of Thyroid Hormone Testing in Human Serum—Results of the Free Thyroxine and Thyrotropin Interlaboratory Comparison Study

Uliana Danilenko, Ashley Ribera, Alicia N. Lyle, Tatiana Buchannan, Hubert W. Vesper, Otoe Sugahara, Li Zhang, Norma Vazquez
article en
8 citations

Abstract

<b><i>Background:</i></b> Performance of thyroid function assays can vary significantly. To address this issue, the Centers for Disease Control and Prevention (CDC) Clinical Standardization Programs conducted an interlaboratory comparison of free thyroxine (fT4) immunoassays (IAs) and laboratory-developed tests (LDTs). This assessment aimed to determine the current performance characteristics of these assays as a first step toward measurement standardization. Thyrotropin (TSH) IAs were also evaluated. <b><i>Methods:</i></b> Assays measured 41 blinded individual-donor sera, including a sample from a pregnant woman (for fT4 analysis only) and three serum pools, with 11.3-32.1 pmol/L (0.881-2.49 ng/dL) fT4 and 0.337-21.6 mIU/L TSH in duplicate over 2 days. Passing-Bablok regression analysis performed pre-recalibration compared assays performance to the CDC fT4 reference measurement procedure (RMP) or TSH all-lab mean (ALM). Additionally, the impact of linear regression-based recalibration of assays to the CDC fT4 RMP or TSH ALM was estimated. Inter-assay agreement of sample classification according to the assay-specific reference interval (RI) was assessed pre- and post-recalibration. <b><i>Results:</i></b> A total of 21 fT4 and 17 TSH assays participated. Pre-recalibration, median biases of TSH measurements to the ALM were -1.2% [confidence interval or CI -1.8% to -0.4%], and good classification agreement among TSH assays was observed. fT4 assays all showed a negative median bias to the RMP, with higher bias among IAs (median: -20.3%, CI [-21.5% to -19.4%]) than LDTs (median: -4.5%, [CI -6.1% to -3.2%]). Of the individual-donor sera, only 21 out of 40 samples were classified uniformly by all fT4 assays, indicating poor inter-assay agreement. Post-recalibration, agreement improved to 33 out of 40 individual-donor sera correctly classified by all tested IAs and LDTs. Similar improvement in post-recalibration median percent bias was observed for fT4 IAs (median: -0.2, [CI -1.2% to 0.6%]) and LDTs (median: -0.3%, [CI -2.5% to 1.4%]). <b><i>Conclusions:</i></b> The comparison among fT4 assays emphasizes the need for measurement standardization to improve accuracy and comparability. This and previous studies demonstrate the possibility to develop common fT4 RIs via standardization, enabling the use of evidence-based clinical guidelines universally in patient care. Recalibration can effectively address high variability in fT4 assays, ensuring consistent diagnostic classification.

ThyroidVol. 35(5)
Centers for Disease Control and Prevention (US)
Zero hunger
Openalex Percentile: Top 2%
Thyroid Disorders and Treatments
10.36
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.