The “16‐gram window” of contact‐force: A new criterion for very high‐power short‐duration ablation
A lower but also a higher mean contact-force was associated with suboptimal vHPSD applications. Hence, a "16-gram window" of contact-force, from 6 to 22 g, could optimize energy application in vHPSD ablation.
Authors
- Michael Böhm (ORCID: https://orcid.org/0000-0002-2976-2514)
- Valérie Pavlicek (ORCID: https://orcid.org/0000-0002-3758-4586)
- Thomas Rostock (ORCID: https://orcid.org/0000-0002-5248-4831)
- Felix Mahfoud (ORCID: https://orcid.org/0000-0002-4425-549X)
- Giuseppe Andrea Ferro (ORCID: https://orcid.org/0000-0002-0622-4203)
- Raphael Spittler (ORCID: https://orcid.org/0000-0001-9941-6302)
- Jan Wintrich (ORCID: https://orcid.org/0000-0001-6518-1453)
- Dimitrios Bismpos
- Alexander P. Benz (ORCID: https://orcid.org/0000-0003-2555-1266)
- Christian Ukena (ORCID: https://orcid.org/0000-0002-9810-4662)
Institutions
- Johannes Gutenberg University Mainz (DE)
- University Hospital of Basel (CH)
- Universitätsklinik Marien Hospital Herne (DE)
- University Medical Center of the Johannes Gutenberg University Mainz (DE)
- Saarland University (DE)
Publication Details
- Journal
- Journal of Arrhythmia
- Published
- 2025-05-06
- DOI
- https://doi.org/10.1002/joa3.70076
- Citations
- 1
- Primary Topic
- Cardiac Arrhythmias and Treatments
- Type
- article
- Field-Weighted Citation Impact
- 1.39