On design, fabrication, and pre-clinical validation of customized 3D-printed dental implant assembly

In the past few decades, 3D-printed dental implants have been manufactured, and significant studies have demonstrated the pre-clinical validation of such systems. However, studies have yet to tackle the ever-present issue of preventing the jumping gap to enhance overall outcomes. The present study details the utilization of patient computed tomography (CT) data to design and subsequently fabricate a multi-component customized dental implant assembly and customized instruments using direct metal laser sintering (DMLS) technology. The workflow was validated for two patient data sets (cases 1 and 2), which were used to render and print custom implant assemblies; the simulation data for these were compared with a commercially available solution. The present study incorporated a prototype stage as well as subjecting the customized implant assemblies to both static (Case 1: 38.89-77.81 MPa vs 75.47-158.09 MPa; Case 2: 83.95-106.65 MPa vs 55.23-126.57 MPa) and dynamic finite element analysis (Case 1: 41.08-84.09 MPa vs 75.45-187.91 MPa; Case 2: 106.81-108.70 MPa vs 79.18-135.48 MPa) along with resonance frequency analysis (Case 1: 7763.2 Hz vs 7003.6 Hz; Case 2: 7910.1 Hz vs 7102.1 Hz) as well as residual stress analysis. The assembly's stress patterns and resonance frequencies were evaluated against a commercially available implant system. It was observed that the customized implant assemblies tended to outperform the commercially available solution in most simulated scenarios.

Authors

Institutions

Publication Details

Journal
Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine
Published
2025-05-01
DOI
https://doi.org/10.1177/09544119251337363
Citations
1
Primary Topic
Dental Implant Techniques and Outcomes
Type
article
Field-Weighted Citation Impact
1.79
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

On design, fabrication, and pre-clinical validation of customized 3D-printed dental implant assembly

Rishab Rishab, Vaibhav Sahni, Smruti Ranjan Pradhan, Ankush Mehta et al.
1 citations
Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine
Dental Implant Techniques and Outcomes
1.79
article

On design, fabrication, and pre-clinical validation of customized 3D-printed dental implant assembly

Rishab Rishab, Vaibhav Sahni, Smruti Ranjan Pradhan, Ankush Mehta, Vishakha Grover, Amrinder Pal Singh, Rupinder Singh, Ashish Jain, Satinder Singh
article en
1 citations

Abstract

In the past few decades, 3D-printed dental implants have been manufactured, and significant studies have demonstrated the pre-clinical validation of such systems. However, studies have yet to tackle the ever-present issue of preventing the jumping gap to enhance overall outcomes. The present study details the utilization of patient computed tomography (CT) data to design and subsequently fabricate a multi-component customized dental implant assembly and customized instruments using direct metal laser sintering (DMLS) technology. The workflow was validated for two patient data sets (cases 1 and 2), which were used to render and print custom implant assemblies; the simulation data for these were compared with a commercially available solution. The present study incorporated a prototype stage as well as subjecting the customized implant assemblies to both static (Case 1: 38.89-77.81 MPa vs 75.47-158.09 MPa; Case 2: 83.95-106.65 MPa vs 55.23-126.57 MPa) and dynamic finite element analysis (Case 1: 41.08-84.09 MPa vs 75.45-187.91 MPa; Case 2: 106.81-108.70 MPa vs 79.18-135.48 MPa) along with resonance frequency analysis (Case 1: 7763.2 Hz vs 7003.6 Hz; Case 2: 7910.1 Hz vs 7102.1 Hz) as well as residual stress analysis. The assembly's stress patterns and resonance frequencies were evaluated against a commercially available implant system. It was observed that the customized implant assemblies tended to outperform the commercially available solution in most simulated scenarios.

Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in MedicineVol. 239(5)
Marwadi University (IN), National Institute of Technical Teachers Training and Research, Chandigarh (IN), Panjab University (IN)
Openalex Percentile: Top 17%
Dental Implant Techniques and Outcomes
1.79
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.