Quantitative Ultrasound Texture Analysis of Breast Tumors: A Comparison of a Cart-Based and a Wireless Ultrasound Scanner

Previous work has demonstrated quantitative ultrasound (QUS) analysis techniques for extracting features and texture features from ultrasound radiofrequency data which can be used to distinguish between benign and malignant breast masses. It is desirable that there be good agreement between estimates of such features acquired using different ultrasound devices. Handheld ultrasound imaging systems are of particular interest as they are compact, relatively inexpensive, and highly portable. This study investigated the agreement between QUS parameters and texture features estimated from clinical ultrasound images of breast tumors acquired using two different ultrasound scanners: a traditional cart-based system and a wireless handheld ultrasound system. The 28 patients who participated were divided into two groups (benign and malignant). The reference phantom technique was used to produce functional estimates of the normalized power spectra and backscatter coefficient for each image. Root mean square differences of feature estimates were calculated for each cohort to quantify the level of feature variation attributable to tissue heterogeneity and differences in system imaging parameters. Cross-system statistical testing using the Mann-Whitney U test was performed on benign and malignant patient cohorts to assess the level of feature estimate agreement between systems, and the Bland-Altman method was employed to assess feature sets for systematic bias introduced by differences in imaging method. The range of <i>p</i>-values was 1.03 × 10<sup>-4</sup> to 0.827 for the benign cohort and 3.03 × 10<sup>-10</sup> to 0.958 for the malignant cohort. For both cohorts, all five of the primary QUS features (MBF, SS, SI, ASD, AAC) were found to be in agreement at the 5% confidence level. A total of 13 of the 20 QUS texture features (65%) were determined to exhibit statistically significant differences in the sample medians of estimates between systems at the 5% confidence level, with the remaining 7 texture features being in agreement. The results showed a comparable magnitude of feature variation between tissue heterogeneity and system effects, as well as a moderate level of statistical agreement between feature sets.

Authors

Institutions

Publication Details

Journal
Journal of Imaging
Published
2025-05-06
DOI
https://doi.org/10.3390/jimaging11050146
Citations
2
Primary Topic
Radiomics and Machine Learning in Medical Imaging
Type
article
Field-Weighted Citation Impact
2.72

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Quantitative Ultrasound Texture Analysis of Breast Tumors: A Comparison of a Cart-Based and a Wireless Ultrasound Scanner

Michael L. Oelze, Lakshmanan Sannachi, Maria Lourdes Anzola Pena, Laurentius O. Osapoetra et al.
2 citations
Journal of Imaging
Radiomics and Machine Learning in Medical Imaging
2.72
article

Quantitative Ultrasound Texture Analysis of Breast Tumors: A Comparison of a Cart-Based and a Wireless Ultrasound Scanner

Michael L. Oelze, Lakshmanan Sannachi, Maria Lourdes Anzola Pena, Laurentius O. Osapoetra, Sonal Gandhi, Joanne Y. W. Yip, Schontal Halstead, Gregory J. Czarnota, Daniel DiCenzo, David Alberico, Frances W. Wright
article en
2 citations

Abstract

Previous work has demonstrated quantitative ultrasound (QUS) analysis techniques for extracting features and texture features from ultrasound radiofrequency data which can be used to distinguish between benign and malignant breast masses. It is desirable that there be good agreement between estimates of such features acquired using different ultrasound devices. Handheld ultrasound imaging systems are of particular interest as they are compact, relatively inexpensive, and highly portable. This study investigated the agreement between QUS parameters and texture features estimated from clinical ultrasound images of breast tumors acquired using two different ultrasound scanners: a traditional cart-based system and a wireless handheld ultrasound system. The 28 patients who participated were divided into two groups (benign and malignant). The reference phantom technique was used to produce functional estimates of the normalized power spectra and backscatter coefficient for each image. Root mean square differences of feature estimates were calculated for each cohort to quantify the level of feature variation attributable to tissue heterogeneity and differences in system imaging parameters. Cross-system statistical testing using the Mann-Whitney U test was performed on benign and malignant patient cohorts to assess the level of feature estimate agreement between systems, and the Bland-Altman method was employed to assess feature sets for systematic bias introduced by differences in imaging method. The range of <i>p</i>-values was 1.03 × 10<sup>-4</sup> to 0.827 for the benign cohort and 3.03 × 10<sup>-10</sup> to 0.958 for the malignant cohort. For both cohorts, all five of the primary QUS features (MBF, SS, SI, ASD, AAC) were found to be in agreement at the 5% confidence level. A total of 13 of the 20 QUS texture features (65%) were determined to exhibit statistically significant differences in the sample medians of estimates between systems at the 5% confidence level, with the remaining 7 texture features being in agreement. The results showed a comparable magnitude of feature variation between tissue heterogeneity and system effects, as well as a moderate level of statistical agreement between feature sets.

Journal of ImagingVol. 11(5)
Sunnybrook Health Science Centre (CA), University of Illinois Urbana-Champaign (US), University of Toronto (CA), Health Sciences Centre (CA), Sunnybrook Research Institute
Natural Sciences and Engineering Research Council of Canada
Good health and well-being
Openalex Percentile: Top 11%
Radiomics and Machine Learning in Medical Imaging
2.72
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.