Dynamics of a Distorted Waveform and Elevated Central Venous Pressure (CVP) Resulting From Coiling of a Central Venous Catheter: A Case Report

Central venous cannulation (CVC) is a routine procedure in critical care and surgical settings. Despite ultrasound guidance, complications like catheter coiling can occur. We report a case of CVC catheter coiling that led to waveform distortion and required corrective measures. A 23-year-old female patient underwent an emergency laparotomy for perforation peritonitis with underlying pulmonary tuberculosis. During the procedure, a triple-lumen CVC was placed in the left internal jugular vein (IJV) after an unsuccessful right IJV cannulation attempt. Upon placement, the CVP was significantly elevated at 31.2 mmHg with a distorted waveform showing exaggerated 'v wave' and diminished 'x descent'. A point-of-care echocardiogram excluded right heart abnormalities. Postoperatively, a chest X-ray revealed coiling of the CVC in the superior vena cava (SVC), forming a fishhook pattern. The catheter was partially withdrawn by 2 cm, resulting in a reduction in CVP to 12 mmHg and normalization of the waveform. The catheter was subsequently removed and replaced without complications. Catheter coiling is an uncommon but significant complication of CVC, even when performed under ultrasound guidance. It can lead to inaccurate CVP readings and waveform distortion, which may affect clinical decisions. Early recognition of abnormal waveforms and partial withdrawal of the catheter can correct the coiling and restore accurate hemodynamic monitoring. This case underscores the importance of waveform analysis as a diagnostic tool for detecting catheter malposition and mechanical complications.

Authors

Institutions

Publication Details

Journal
Cureus
Published
2025-05-06
DOI
https://doi.org/10.7759/cureus.83561
Primary Topic
Central Venous Catheters and Hemodialysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Dynamics of a Distorted Waveform and Elevated Central Venous Pressure (CVP) Resulting From Coiling of a Central Venous Catheter: A Case Report

Sagar Jolly, Shashank Paliwal, Hemanthkumar Tamilchelvan, Jyoti Sharma et al.
4 links
Cureus
Central Venous Catheters and Hemodialysis
article

Dynamics of a Distorted Waveform and Elevated Central Venous Pressure (CVP) Resulting From Coiling of a Central Venous Catheter: A Case Report

Sagar Jolly, Shashank Paliwal, Hemanthkumar Tamilchelvan, Jyoti Sharma, Navneh Samagh
article en

Abstract

Central venous cannulation (CVC) is a routine procedure in critical care and surgical settings. Despite ultrasound guidance, complications like catheter coiling can occur. We report a case of CVC catheter coiling that led to waveform distortion and required corrective measures. A 23-year-old female patient underwent an emergency laparotomy for perforation peritonitis with underlying pulmonary tuberculosis. During the procedure, a triple-lumen CVC was placed in the left internal jugular vein (IJV) after an unsuccessful right IJV cannulation attempt. Upon placement, the CVP was significantly elevated at 31.2 mmHg with a distorted waveform showing exaggerated 'v wave' and diminished 'x descent'. A point-of-care echocardiogram excluded right heart abnormalities. Postoperatively, a chest X-ray revealed coiling of the CVC in the superior vena cava (SVC), forming a fishhook pattern. The catheter was partially withdrawn by 2 cm, resulting in a reduction in CVP to 12 mmHg and normalization of the waveform. The catheter was subsequently removed and replaced without complications. Catheter coiling is an uncommon but significant complication of CVC, even when performed under ultrasound guidance. It can lead to inaccurate CVP readings and waveform distortion, which may affect clinical decisions. Early recognition of abnormal waveforms and partial withdrawal of the catheter can correct the coiling and restore accurate hemodynamic monitoring. This case underscores the importance of waveform analysis as a diagnostic tool for detecting catheter malposition and mechanical complications.

CureusVol. 17(5)
Central University of Punjab (IN), Metropolitan Hospital Center (US)
Openalex Percentile: Top 84%
Central Venous Catheters and Hemodialysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Dynamics of a Distorted Waveform and Elevated Central Venous Pressure (CVP) Resulting From Coiling of a Central Venous Catheter: A Case Report — Sagar Jolly, Shashank Paliwal, et al. · Cureus (2025) | TGRS Research Map | TGRS