Optimizing Input Feature Sets Using Catch-22 and Personalization for an Accurate and Reliable Estimation of Continuous, Cuffless Blood Pressure

Nocturnal monitoring of continuous, cuffless blood pressure (BP) can unleash a whole new world for the prognostication of cardiovascular and other diseases due to its strong predictive capability. Nevertheless, the lack of an accurate and reliable method, primarily due to confounding variables, has prevented its widespread clinical adoption. Herein, we demonstrate how optimized machine learning using the Catch-22 features, when applied to the photoplethysmogram waveform and personalized with direct BP data through transfer learning, can accurately estimate systolic and diastolic BP. After training with a hemodynamically compromised VitalDB "calibration-free" dataset (n = 1293), the systolic and diastolic BP tested on a distinct VitalDB dataset that met AAMI criteria (n = 116) had acceptable error biases of -1.85 mm Hg and 0.11 mm Hg, respectively [within the 5 mm Hg IEC/ANSI/AAMI 80601-2-30, 2018 standard], but standard deviation (SD) errors of 19.55 mm Hg and 11.55 mm Hg, respectively [exceeding the stipulated 8 mm Hg limit]. However, personalization using an initial calibration data segment and subsequent use of transfer learning to fine-tune the pretrained model produced acceptable mean (-1.31 mm Hg and 0.10 mm Hg) and SD (7.91 mm Hg and 4.59 mm Hg) errors for systolic and diastolic BP, respectively. Levene's test for variance found that the personalization method significantly outperformed (<i>p</i> < 0.05) the calibration-free method, but there was no difference between three machine learning methods. Optimized multimodal Catch-22 features, coupled with personalization, demonstrate great promise in the clinical adoption of continuous, cuffless blood pressure estimation in applications such as nocturnal BP monitoring.

Authors

Institutions

Publication Details

Journal
Bioengineering
Published
2025-05-06
DOI
https://doi.org/10.3390/bioengineering12050493
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Optimizing Input Feature Sets Using Catch-22 and Personalization for an Accurate and Reliable Estimation of Continuous, Cuffless Blood Pressure

Rajesh S. Kasbekar, Anita Goel, Edward A. Clancy, Songbai Ji et al.
Bioengineering
Non-Invasive Vital Sign Monitoring
article

Optimizing Input Feature Sets Using Catch-22 and Personalization for an Accurate and Reliable Estimation of Continuous, Cuffless Blood Pressure

Rajesh S. Kasbekar, Anita Goel, Edward A. Clancy, Songbai Ji, Srinivasan Radhakrishnan
article en

Abstract

Nocturnal monitoring of continuous, cuffless blood pressure (BP) can unleash a whole new world for the prognostication of cardiovascular and other diseases due to its strong predictive capability. Nevertheless, the lack of an accurate and reliable method, primarily due to confounding variables, has prevented its widespread clinical adoption. Herein, we demonstrate how optimized machine learning using the Catch-22 features, when applied to the photoplethysmogram waveform and personalized with direct BP data through transfer learning, can accurately estimate systolic and diastolic BP. After training with a hemodynamically compromised VitalDB "calibration-free" dataset (n = 1293), the systolic and diastolic BP tested on a distinct VitalDB dataset that met AAMI criteria (n = 116) had acceptable error biases of -1.85 mm Hg and 0.11 mm Hg, respectively [within the 5 mm Hg IEC/ANSI/AAMI 80601-2-30, 2018 standard], but standard deviation (SD) errors of 19.55 mm Hg and 11.55 mm Hg, respectively [exceeding the stipulated 8 mm Hg limit]. However, personalization using an initial calibration data segment and subsequent use of transfer learning to fine-tune the pretrained model produced acceptable mean (-1.31 mm Hg and 0.10 mm Hg) and SD (7.91 mm Hg and 4.59 mm Hg) errors for systolic and diastolic BP, respectively. Levene's test for variance found that the personalization method significantly outperformed (<i>p</i> < 0.05) the calibration-free method, but there was no difference between three machine learning methods. Optimized multimodal Catch-22 features, coupled with personalization, demonstrate great promise in the clinical adoption of continuous, cuffless blood pressure estimation in applications such as nocturnal BP monitoring.

BioengineeringVol. 12(5)
Worcester Polytechnic Institute (US), Northeastern University (US), Harvard University (US)
Good health and well-being
Openalex Percentile: Top 91%
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.