Mechanical and in Vitro Testing of a Biomechanically Fidelic, Compliant Mechanism Based, Motion-Preserving Lumbar Interbody Device

Abstract A common approach to resolving discogenic low back pain involves replacing one or more degenerated spinal discs with a total disc replacement device. However, existing solutions for intervertebral disc replacement are unable to fully capture the kinetic and kinematic characteristics (i.e., the quality of motion) of the intact spinal disc. In the present work, a novel single-piece compliant mechanism driven, motion-preserving lumbar intervertebral implant design is described. Prototypes were manufactured from Ti6Al4V and evaluated using benchtop mechanical and in vitro biomechanical testing. ASTM F2346 static testing procedures were followed to assess the design's compressive, shear, and torsional properties. Similarly, the forces necessary to cause the device to be ejected from the interbody space and the force required to cause subsidence were tested. in vitro testing was conducted with fresh-frozen human cadaveric lumbar spinal segments to analyze the quality of motion of the intact segments and after they were instrumented with the compliant interbody device prototype. The design was robust in static compressive, shear, and torsional loading. Expulsion and subsidence test results were comparable to devices currently in use. in vitro testing indicated that when appropriately placed in the intervertebral space, the instrumented segment's quality of motion closely replicated the intact segment.

Authors

Institutions

Publication Details

Journal
Journal of Medical Devices
Published
2025-05-06
DOI
https://doi.org/10.1115/1.4068611
Primary Topic
Spine and Intervertebral Disc Pathology
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Mechanical and in Vitro Testing of a Biomechanically Fidelic, Compliant Mechanism Based, Motion-Preserving Lumbar Interbody Device

Anton E. Bowden, Tyler Peterson, Daniel J Orr, Ron Anderson et al.
Journal of Medical Devices
Spine and Intervertebral Disc Pathology
article

Mechanical and in Vitro Testing of a Biomechanically Fidelic, Compliant Mechanism Based, Motion-Preserving Lumbar Interbody Device

Anton E. Bowden, Tyler Peterson, Daniel J Orr, Ron Anderson, Anna Jensen, Andrew Thomson
article en

Abstract

Abstract A common approach to resolving discogenic low back pain involves replacing one or more degenerated spinal discs with a total disc replacement device. However, existing solutions for intervertebral disc replacement are unable to fully capture the kinetic and kinematic characteristics (i.e., the quality of motion) of the intact spinal disc. In the present work, a novel single-piece compliant mechanism driven, motion-preserving lumbar intervertebral implant design is described. Prototypes were manufactured from Ti6Al4V and evaluated using benchtop mechanical and in vitro biomechanical testing. ASTM F2346 static testing procedures were followed to assess the design's compressive, shear, and torsional properties. Similarly, the forces necessary to cause the device to be ejected from the interbody space and the force required to cause subsidence were tested. in vitro testing was conducted with fresh-frozen human cadaveric lumbar spinal segments to analyze the quality of motion of the intact segments and after they were instrumented with the compliant interbody device prototype. The design was robust in static compressive, shear, and torsional loading. Expulsion and subsidence test results were comparable to devices currently in use. in vitro testing indicated that when appropriately placed in the intervertebral space, the instrumented segment's quality of motion closely replicated the intact segment.

Journal of Medical DevicesVol. 19(3)
Brigham Young University (US)
Openalex Percentile: Top 86%
Spine and Intervertebral Disc Pathology
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Mechanical and in Vitro Testing of a Biomechanically Fidelic, Compliant Mechanism Based, Motion-Preserving Lumbar Interbody Device — Anton E. Bowden, Tyler Peterson, et al. · Journal of Medical Devices (2025) | TGRS Research Map | TGRS