A performance evaluation of silver nanorods PDMS flexible dry electrodes for electrocardiogram monitoring

Extensive research is being conducted in fabricating flexible dry electrodes for electrocardiogram monitoring, but the electrodes' efficacy in clinical settings remains underexplored. In transition from research to commercial settings, investigating the electrode's performance in real-time monitoring and patient's comfort is very crucial. This study compares the ECG signal quality between flexible silver nanorods embedded in polydimethylsiloxane (AgNRs-PDMS) dry electrodes and commercially available metal electrodes. This study, conducted in a hospital, involves 50 subjects (40 males, 10 females; age range: 20-74) among which 41 were with cardiovascular disease and 9 normal subjects. The fabricated dry electrodes are biocompatible and have a lower skin-to-electrode impedance than the commercial electrodes, resulting in high-fidelity ECG signals. Signal quality was assessed based on parameters such as signal-to-noise ratio, mean amplitude, maximum amplitude, power spectral density, and heart rate comparison. The AgNRs-PDMS electrodes demonstrated superior SNR, confirmed using a paired t-test, with a p-value close to 0, indicating a significant difference in comparison with commercial electrodes. The amplitude of ECG signals captured by AgNRs-PDMS electrodes and the heart rate were observed to be comparable to metal electrodes. For automated arrhythmia classification of the ECG signals, two models were implemented. The first model utilized R-R interval for arrhythmic rhythm classification, while the second model used principal component analysis (PCA) for dimensionality reduction followed by support vector machine (SVM) to classify arrhythmic beats. Large arrhythmia data sets like the MIT-BIH arrhythmia database were used for training and validating the above models. Accuracy results from the MIT-BIH test data set were 97% for the R-R interval method and 93% for the SVM method. The heart beats obtained from an arrhythmic patient using commercial metal electrodes and AgNRs-PDMS electrodes were classified using the classifiers. The AgNRs-PDMS dry electrodes offer superior signal quality, ease of use due to gel-free nature, and reusability, making them a promising alternative to commercial electrodes for clinical ECG monitoring.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2025-05-06
DOI
https://doi.org/10.1038/s41598-025-95057-z
Citations
5
Primary Topic
ECG Monitoring and Analysis
Type
article
Field-Weighted Citation Impact
6.94

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A performance evaluation of silver nanorods PDMS flexible dry electrodes for electrocardiogram monitoring

Ghanshyam Kumar, C. M. Vidhya, Bhanu Duggal, J. P. Singh et al.
5 citations
Scientific Reports
ECG Monitoring and Analysis
6.94
article

A performance evaluation of silver nanorods PDMS flexible dry electrodes for electrocardiogram monitoring

Ghanshyam Kumar, C. M. Vidhya, Bhanu Duggal, J. P. Singh, Yogita Maithani
article en
5 citations

Abstract

Extensive research is being conducted in fabricating flexible dry electrodes for electrocardiogram monitoring, but the electrodes' efficacy in clinical settings remains underexplored. In transition from research to commercial settings, investigating the electrode's performance in real-time monitoring and patient's comfort is very crucial. This study compares the ECG signal quality between flexible silver nanorods embedded in polydimethylsiloxane (AgNRs-PDMS) dry electrodes and commercially available metal electrodes. This study, conducted in a hospital, involves 50 subjects (40 males, 10 females; age range: 20-74) among which 41 were with cardiovascular disease and 9 normal subjects. The fabricated dry electrodes are biocompatible and have a lower skin-to-electrode impedance than the commercial electrodes, resulting in high-fidelity ECG signals. Signal quality was assessed based on parameters such as signal-to-noise ratio, mean amplitude, maximum amplitude, power spectral density, and heart rate comparison. The AgNRs-PDMS electrodes demonstrated superior SNR, confirmed using a paired t-test, with a p-value close to 0, indicating a significant difference in comparison with commercial electrodes. The amplitude of ECG signals captured by AgNRs-PDMS electrodes and the heart rate were observed to be comparable to metal electrodes. For automated arrhythmia classification of the ECG signals, two models were implemented. The first model utilized R-R interval for arrhythmic rhythm classification, while the second model used principal component analysis (PCA) for dimensionality reduction followed by support vector machine (SVM) to classify arrhythmic beats. Large arrhythmia data sets like the MIT-BIH arrhythmia database were used for training and validating the above models. Accuracy results from the MIT-BIH test data set were 97% for the R-R interval method and 93% for the SVM method. The heart beats obtained from an arrhythmic patient using commercial metal electrodes and AgNRs-PDMS electrodes were classified using the classifiers. The AgNRs-PDMS dry electrodes offer superior signal quality, ease of use due to gel-free nature, and reusability, making them a promising alternative to commercial electrodes for clinical ECG monitoring.

Scientific ReportsVol. 15(1)
All India Institute of Medical Sciences Rishikesh (IN), Indian Institute of Technology Delhi (IN)
Ministry of Education, India, All-India Institute of Medical Sciences, Indian Institute of Technology Delhi, Biotechnology Industry Research Assistance Council, Division of Human Resource Development
Openalex Percentile: Top 3%
ECG Monitoring and Analysis
6.94
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.