The stress distribution of implants made of Polyaryletherketones: A 3D Finite Element Analysis

OBJECTIVE: To explore the stress distribution on the bone-implant structures caused by the Polyetheretherketone (PEEK), Polyetherketoneketone (PEKK), and titanium fixture/abutments by using the three-dimensional (3D) finite element analysis (FEA). MATERIALS AND METHODS: Six models composed of titanium, PEEK, and PEKK implant/abutments under vertical (250 N) and 45˚ oblique (100 N) loading were studied. The obtained principal and von Mises stress values from the models were evaluated. RESULTS: von Mises stresses were found to be highest in titanium implants and abutments under vertical and oblique loads. Extremely increased stress values were observed in the screws of the Polyaryletherketone (PAEK) models compared to titanium models. Lower principal stresses were observed in titanium models than PAEK models in the cancellous and cortical bone under vertical and oblique loads. CONCLUSION: PAEKs transmitted more stress to the peri-implant bone. Stress distribution in titanium models were more homogenous while stress concentrated in the bone adjacent to the coronal part of the implants and neck of the implants in PAEKs.

Authors

Institutions

Publication Details

Journal
Acta Odontologica Turcica
Published
2025-05-05
DOI
https://doi.org/10.17214/gaziaot.1420859
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The stress distribution of implants made of Polyaryletherketones: A 3D Finite Element Analysis

Ahmet Serkan Küçükekenci, Muhammed Latif Bekçi, Mehmet Ali Güler
4 links
Acta Odontologica Turcica
Orthopaedic implants and arthroplasty
article

The stress distribution of implants made of Polyaryletherketones: A 3D Finite Element Analysis

Ahmet Serkan Küçükekenci, Muhammed Latif Bekçi, Mehmet Ali Güler
article en

Abstract

OBJECTIVE: To explore the stress distribution on the bone-implant structures caused by the Polyetheretherketone (PEEK), Polyetherketoneketone (PEKK), and titanium fixture/abutments by using the three-dimensional (3D) finite element analysis (FEA). MATERIALS AND METHODS: Six models composed of titanium, PEEK, and PEKK implant/abutments under vertical (250 N) and 45˚ oblique (100 N) loading were studied. The obtained principal and von Mises stress values from the models were evaluated. RESULTS: von Mises stresses were found to be highest in titanium implants and abutments under vertical and oblique loads. Extremely increased stress values were observed in the screws of the Polyaryletherketone (PAEK) models compared to titanium models. Lower principal stresses were observed in titanium models than PAEK models in the cancellous and cortical bone under vertical and oblique loads. CONCLUSION: PAEKs transmitted more stress to the peri-implant bone. Stress distribution in titanium models were more homogenous while stress concentrated in the bone adjacent to the coronal part of the implants and neck of the implants in PAEKs.

Acta Odontologica TurcicaVol. 42(2)
Ordu University (TR)
Openalex Percentile: Top 83%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

The stress distribution of implants made of Polyaryletherketones: A 3D Finite Element Analysis — Ahmet Serkan Küçükekenci, Muhammed Latif Bekçi, et al. · Acta Odontologica Turcica (2025) | TGRS Research Map | TGRS