Design and simulation of a 7.0 T conduction cooled superconducting magnet

A Conduction cooled superconducting magnet (SM) for human magnetic resonance imaging, made of Nb<sub>3</sub>Sn superconducting coils, has been designed. The magnet features a warm bore with a diameter of 850 mm and a central field strength of 7 T. The size and positioning of a bundle of seven epoxy-impregnated coils were determined using both linear and nonlinear optimization methods. These methods ensured that the superconducting coils achieved a magnetic field uniformity of 10 ppm within a 400 mm diameter spherical volume (DSV). The magnet was designed to operate at a current of 250 A in a cryogenic system, and its temperature will be maintained below 8 K through conduction cooling provided by a Gifford-McMahon (G-M) cryocooler at its second-stage cooling capacity. The SM and its cryogenic system were designed according to a comprehensive analysis of the mechanical and thermal features of the magnet system. The design was evaluated and validated by Finite Elements simulations. The results confirmed that the magnet is capable of stable and reliable operation at low temperatures, effectively achieving conduction cooling.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2025-05-05
DOI
https://doi.org/10.1038/s41598-025-00643-w
Citations
3
Primary Topic
Superconducting Materials and Applications
Type
article
Field-Weighted Citation Impact
2.02

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Design and simulation of a 7.0 T conduction cooled superconducting magnet

Zhichao Feng, Junsheng Cheng, Shuo Wang, Xu Zhao et al.
3 citations
Scientific Reports
Superconducting Materials and Applications
2.02
article

Design and simulation of a 7.0 T conduction cooled superconducting magnet

Zhichao Feng, Junsheng Cheng, Shuo Wang, Xu Zhao, Qiuliang Wang, Qi Chen, Jianhua Liu, Lei Wang, Hui Wang, Yaohui Wang, Jing Xiao, Ce Xu, Huafang Wu
article en
3 citations

Abstract

A Conduction cooled superconducting magnet (SM) for human magnetic resonance imaging, made of Nb<sub>3</sub>Sn superconducting coils, has been designed. The magnet features a warm bore with a diameter of 850 mm and a central field strength of 7 T. The size and positioning of a bundle of seven epoxy-impregnated coils were determined using both linear and nonlinear optimization methods. These methods ensured that the superconducting coils achieved a magnetic field uniformity of 10 ppm within a 400 mm diameter spherical volume (DSV). The magnet was designed to operate at a current of 250 A in a cryogenic system, and its temperature will be maintained below 8 K through conduction cooling provided by a Gifford-McMahon (G-M) cryocooler at its second-stage cooling capacity. The SM and its cryogenic system were designed according to a comprehensive analysis of the mechanical and thermal features of the magnet system. The design was evaluated and validated by Finite Elements simulations. The results confirmed that the magnet is capable of stable and reliable operation at low temperatures, effectively achieving conduction cooling.

Scientific ReportsVol. 15(1)
Qilu University of Technology (CN), Chinese Academy of Sciences (CN), Institute of Electrical Engineering (CN), Shandong Academy of Sciences (CN)
National Key Research and Development Program of China
Affordable and clean energy
Openalex Percentile: Top 14%
Superconducting Materials and Applications
2.02
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.