Construction and characterization of inexpensive PRT differential thermometer in temperature interval 20–100 °C

A precise, inexpensive, and mercury-free differential platinum resistance thermometer based on the widely known and used 4 × Pt1000 Wheatstone bridge scheme has been constructed and tested for direct temperature difference measurements. Both a theoretical overview and an experimental study of the 4 × Pt1000 differential thermometer were conducted in laboratory conditions at a temperature interval of 20-100 °C. A comparison between the parameters of the 4 × Pt1000 measurement instrument and those of an alternative less accurate 2 × Pt1000 differential thermometer was made. The results show that the 4 × Pt1000 thermometer performs significantly better in terms of stability and accuracy. A maximal relative error below 0.9% has been achieved with the 4 × Pt1000 measurement device, which equals 0.20 °C in absolute units. Such accuracy is considered higher than that of some reference thermocouples, type S (0.5-1.0 °C). The presented 4 × Pt1000 differential thermometer is a low-cost substitute for thermocouples in various precise narrow-margin measurements of temperature differences in the fields of physics, metrology, chemistry, etc.

Authors

Institutions

Publication Details

Journal
Review of Scientific Instruments
Published
2025-05-01
DOI
https://doi.org/10.1063/5.0257253
Primary Topic
Calibration and Measurement Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Construction and characterization of inexpensive PRT differential thermometer in temperature interval 20–100 °C

Ivo I. Lozanov, L. Lyutov
Review of Scientific Instruments
Calibration and Measurement Techniques
article

Construction and characterization of inexpensive PRT differential thermometer in temperature interval 20–100 °C

Ivo I. Lozanov, L. Lyutov
article en

Abstract

A precise, inexpensive, and mercury-free differential platinum resistance thermometer based on the widely known and used 4 × Pt1000 Wheatstone bridge scheme has been constructed and tested for direct temperature difference measurements. Both a theoretical overview and an experimental study of the 4 × Pt1000 differential thermometer were conducted in laboratory conditions at a temperature interval of 20-100 °C. A comparison between the parameters of the 4 × Pt1000 measurement instrument and those of an alternative less accurate 2 × Pt1000 differential thermometer was made. The results show that the 4 × Pt1000 thermometer performs significantly better in terms of stability and accuracy. A maximal relative error below 0.9% has been achieved with the 4 × Pt1000 measurement device, which equals 0.20 °C in absolute units. Such accuracy is considered higher than that of some reference thermocouples, type S (0.5-1.0 °C). The presented 4 × Pt1000 differential thermometer is a low-cost substitute for thermocouples in various precise narrow-margin measurements of temperature differences in the fields of physics, metrology, chemistry, etc.

Review of Scientific InstrumentsVol. 96(5)
Bulgarian Academy of Sciences (BG), Institute of Metal Science, Equipment and Technologies (BG), Sofia University "St. Kliment Ohridski" (BG)
Openalex Percentile: Top 87%
Calibration and Measurement Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.