Contamination of microbubbles of air may occur at all investigated measurement points during hemodialysis

Microbubbles (MBs) of air occur in the hemodialysis (HD) extracorporeal circuit and may enter the bloodlines of the patient. The aim of the present study was to investigate possible sites of contamination. Seventeen patients performed 20 HD (Baxter AK200S <i>n</i> = 5 and Artis <i>n</i> = 15) and 930 ultrasound measurements of MBs/min (Hatteland CMD10 device). Detection ranges were diameters between 2.5 and 50 µm. Hemodiafiltration with postdilution (HDF-post) was performed in 14 dialyses, predilution (HDF-pre) in 1 dialysis, and HD using hemocontrol (HDhc) in 5 dialyses. Measurement points were M1-after the blood access, M2-before the dialyzer, M3-after the dialyzer, and M4-after the venous chamber. At each point, 10 measures of MBs were performed. MB contamination of the blood was larger at all points when the access was an arteriovenous fistula compared to a central dialysis catheter (<i>p</i> < 0.001). MB levels with the AK200 versus the Artis were lower at M1, higher at M2 (<i>p</i> ⩽ 0.005), and were similar at M3 and M4. HDF-pre had fewer MBs than HDF-post, whereas HDhc had more MBs than HDF-post (<i>p</i> < 0.001). An increase of MBs was seen at M2 during an internal "Autotest." No air alarms were induced during dialyses. MBs were detected in the extra corporeal circuit at all points investigated. The venous chambers used did not significantly reduce contamination. The detected MBs did not induce air alarms when the blood returned to the patient.

Authors

Institutions

Publication Details

Journal
The International Journal of Artificial Organs
Published
2025-05-01
DOI
https://doi.org/10.1177/03913988251334953
Primary Topic
Central Venous Catheters and Hemodialysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Contamination of microbubbles of air may occur at all investigated measurement points during hemodialysis

Bernd Stegmayr, Per Jönsson
The International Journal of Artificial Organs
Central Venous Catheters and Hemodialysis
article

Contamination of microbubbles of air may occur at all investigated measurement points during hemodialysis

Bernd Stegmayr, Per Jönsson
article en

Abstract

Microbubbles (MBs) of air occur in the hemodialysis (HD) extracorporeal circuit and may enter the bloodlines of the patient. The aim of the present study was to investigate possible sites of contamination. Seventeen patients performed 20 HD (Baxter AK200S <i>n</i> = 5 and Artis <i>n</i> = 15) and 930 ultrasound measurements of MBs/min (Hatteland CMD10 device). Detection ranges were diameters between 2.5 and 50 µm. Hemodiafiltration with postdilution (HDF-post) was performed in 14 dialyses, predilution (HDF-pre) in 1 dialysis, and HD using hemocontrol (HDhc) in 5 dialyses. Measurement points were M1-after the blood access, M2-before the dialyzer, M3-after the dialyzer, and M4-after the venous chamber. At each point, 10 measures of MBs were performed. MB contamination of the blood was larger at all points when the access was an arteriovenous fistula compared to a central dialysis catheter (<i>p</i> < 0.001). MB levels with the AK200 versus the Artis were lower at M1, higher at M2 (<i>p</i> ⩽ 0.005), and were similar at M3 and M4. HDF-pre had fewer MBs than HDF-post, whereas HDhc had more MBs than HDF-post (<i>p</i> < 0.001). An increase of MBs was seen at M2 during an internal "Autotest." No air alarms were induced during dialyses. MBs were detected in the extra corporeal circuit at all points investigated. The venous chambers used did not significantly reduce contamination. The detected MBs did not induce air alarms when the blood returned to the patient.

The International Journal of Artificial OrgansVol. 48(5)
Umeå University (SE)
Openalex Percentile: Top 84%
Central Venous Catheters and Hemodialysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.