A Potential Pitfall in the Interpretation of Microscope-Integrated Fluorescence Angiography: The Center–Periphery Effect
Peripheral signal loss in microscope-integrated ICG-FA must be considered during clinical interpretation and for the quantification of tissue perfusion. In clinical practice during reconstructive free-flap surgery, a 30 cm WD, 3.5 MF, and manually opened light beam diameter should be applied to achieve the most homogeneous image.