Scheitelbrechwertprofile weicher rotationssymmetrischer Kontaktlinsen in Abhängigkeit des Aperturdurchmessers

Purpose. The aim of the study was to measure and to compare the power profiles of market-leading single vision and multifocal soft contact lenses. Material and Methods. A high-resolution instrument based on the phase-shifting schlieren method (NIMO TR1504, Lambda-X, Belgium) was used to measure the optical power distribution of single vision and multifocal contact lenses with −3.00 D, −6.00 D, −9.00 D and +3.00 D. The optical power of the single vision lenses was measured at six different aperture diameters ranging from 2.0 mm to 7.0 mm in steps of 1.0 mm. Each multifocal design was measured with an aperture of 7.0 mm. A total of 200 contact lenses was measured, five lenses per power and type. Results. All single vision contact lenses with negative back vertex power showed a negative increase in lens power with increased distance from centre. The changes in power between the central and peripheral aperture were statistically and clinically significant for every negative refractive power and type (p < 0.05). Etafilcon A showed the highest negative increase with greater apertures with values up to 1.19 ± 0.04 D. Lotrafilcon A showed overall the least deviation with maximum values of 0.61 ± 0.05 D. Each type of multifocal contact lens exhibited different power profiles. The profiles were very consistent across all negative back vertex powers. Conclusion. As expected, the negative single vision contact lenses exhibited an increase in lens power towards the periphery due to spherical aberration. Greater aberration resulted in a higher negative deviation. The analysis of power profiles of multifocal lenses provides valuable information for fitting lenses for presbyopes considering their pupil size.

Authors

Publication Details

Journal
Optometry & contact lenses
Published
2025-04-29
DOI
https://doi.org/10.54352/dozv.gsss2628
Primary Topic
Tribology and Lubrication Engineering
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Scheitelbrechwertprofile weicher rotationssymmetrischer Kontaktlinsen in Abhängigkeit des Aperturdurchmessers

Sebastian Marx, Jessica Kemper, Sebastian Schubert, Wolfgang Sickenberger
Optometry & contact lenses
Tribology and Lubrication Engineering
article

Scheitelbrechwertprofile weicher rotationssymmetrischer Kontaktlinsen in Abhängigkeit des Aperturdurchmessers

Sebastian Marx, Jessica Kemper, Sebastian Schubert, Wolfgang Sickenberger
article en

Abstract

Purpose. The aim of the study was to measure and to compare the power profiles of market-leading single vision and multifocal soft contact lenses. Material and Methods. A high-resolution instrument based on the phase-shifting schlieren method (NIMO TR1504, Lambda-X, Belgium) was used to measure the optical power distribution of single vision and multifocal contact lenses with −3.00 D, −6.00 D, −9.00 D and +3.00 D. The optical power of the single vision lenses was measured at six different aperture diameters ranging from 2.0 mm to 7.0 mm in steps of 1.0 mm. Each multifocal design was measured with an aperture of 7.0 mm. A total of 200 contact lenses was measured, five lenses per power and type. Results. All single vision contact lenses with negative back vertex power showed a negative increase in lens power with increased distance from centre. The changes in power between the central and peripheral aperture were statistically and clinically significant for every negative refractive power and type (p < 0.05). Etafilcon A showed the highest negative increase with greater apertures with values up to 1.19 ± 0.04 D. Lotrafilcon A showed overall the least deviation with maximum values of 0.61 ± 0.05 D. Each type of multifocal contact lens exhibited different power profiles. The profiles were very consistent across all negative back vertex powers. Conclusion. As expected, the negative single vision contact lenses exhibited an increase in lens power towards the periphery due to spherical aberration. Greater aberration resulted in a higher negative deviation. The analysis of power profiles of multifocal lenses provides valuable information for fitting lenses for presbyopes considering their pupil size.

Optometry & contact lensesVol. 5(4)
Openalex Percentile: Top 91%
Tribology and Lubrication Engineering
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.