To evaluate the effects of lubricating eye drops on the repeatability of keratometry (K) measurements in patients presenting for cataract surgery

Purpose – To evaluate the impact of lubricating eye drops on the repeatability of keratometry (K) measurements in cataract surgery patients. Design – Prospective, comparative, hospital-based study. Method – Keratometry was performed using an autorefractometer. At Visit 1, baseline K readings were taken twice before instilling drops. After instillation, K readings were recorded at 30 seconds and 2 minutes. Tear Break-Up Time (TBUT) was measured to classify patients into normal and dry eye groups. A second set of readings was taken the following morning. At Visit 2, one week postoperatively, a third set of K1 and K2 readings was recorded. Result – Variations in K1 and K2 readings were analyzed across time points and between tear film groups. In the dry eye group, significant changes in K1 were observed across intervals (p < 0.0001). Significant differences in K1 were also found between normal and dry eyes at 30 seconds and 2 minutes post-instillation. For K2, notable differences occurred in both groups, particularly between pre- and postoperative readings and all intervals in the dry eye group. Conclusion – Lubricating drops significantly affect keratometry, especially in dry eyes. Waiting over 5 minutes post-instillation is advised before taking measurements.

Authors

Institutions

Publication Details

Journal
International Journal of Health Sciences
Published
2025-05-04
DOI
https://doi.org/10.53730/ijhs.v9n2.15634
Primary Topic
Ocular Surface and Contact Lens
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

To evaluate the effects of lubricating eye drops on the repeatability of keratometry (K) measurements in patients presenting for cataract surgery

Arvind L. Tenagi, Dhruv Goyal, Dr B K Bhagyajyothi, Dr.Krutika Thorat et al.
International Journal of Health Sciences
Ocular Surface and Contact Lens
article

To evaluate the effects of lubricating eye drops on the repeatability of keratometry (K) measurements in patients presenting for cataract surgery

Arvind L. Tenagi, Dhruv Goyal, Dr B K Bhagyajyothi, Dr.Krutika Thorat, Dr Mitali Mangoli
article en

Abstract

Purpose – To evaluate the impact of lubricating eye drops on the repeatability of keratometry (K) measurements in cataract surgery patients. Design – Prospective, comparative, hospital-based study. Method – Keratometry was performed using an autorefractometer. At Visit 1, baseline K readings were taken twice before instilling drops. After instillation, K readings were recorded at 30 seconds and 2 minutes. Tear Break-Up Time (TBUT) was measured to classify patients into normal and dry eye groups. A second set of readings was taken the following morning. At Visit 2, one week postoperatively, a third set of K1 and K2 readings was recorded. Result – Variations in K1 and K2 readings were analyzed across time points and between tear film groups. In the dry eye group, significant changes in K1 were observed across intervals (p < 0.0001). Significant differences in K1 were also found between normal and dry eyes at 30 seconds and 2 minutes post-instillation. For K2, notable differences occurred in both groups, particularly between pre- and postoperative readings and all intervals in the dry eye group. Conclusion – Lubricating drops significantly affect keratometry, especially in dry eyes. Waiting over 5 minutes post-instillation is advised before taking measurements.

International Journal of Health SciencesVol. 9(2)
Jawaharlal Nehru Medical College (IN)
Openalex Percentile: Top 87%
Ocular Surface and Contact Lens
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

To evaluate the effects of lubricating eye drops on the repeatability of keratometry (K) measurements in patients presenting for cataract surgery — Arvind L. Tenagi, Dhruv Goyal, et al. · International Journal of Health Sciences (2025) | TGRS Research Map | TGRS