Simulation and analysis of non-navigational errors in robot-assisted pedicle Kirschner wire placement surgery
Operator habits and vertebral fixation stiffness are the primary factors influencing non-navigational errors, while guide-to-bone surface distance and robotic arm stiffness are secondary factors. This study recommends adopting Habit 1 in clinical surgeries, minimizing the guide-to-bone surface distance, and enhancing the stiffness of the robotic arm and vertebral fixation to reduce non-navigational errors and improve the accuracy of robot-assisted pedicle screw placement.
Authors
- Boyao Wang
- Yongkang Yang (ORCID: https://orcid.org/0000-0002-1021-7612)
- Chang Liu (ORCID: https://orcid.org/0000-0002-9671-5793)
- Liang Li (ORCID: https://orcid.org/0000-0002-0417-8995)
- Yanan Jia (ORCID: https://orcid.org/0009-0008-8883-5913)
Institutions
- Nanjing Medical University (CN)
Publication Details
- Journal
- Journal of Orthopaedic Surgery and Research
- Published
- 2025-05-02
- DOI
- https://doi.org/10.1186/s13018-025-05790-4
- Primary Topic
- Spinal Fractures and Fixation Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Government of Jiangsu Province
- Natural Science Foundation of Jiangsu Province
- Nanjing Medical University