Development and Validation of an LC-MS/MS Method for the Determination of Lamivudine, Tenofovir & Efavirenz in human plasma

Introduction: Therapeutic drug monitoring of Lamivudine, tenofovir and efavirenz, three commonly used antiretroviral drugs, is important to maximize effectiveness while minimizing side effects. Materials and Methods: Chromatographic separation was done on Xterra, C18 (2), 150 X 4.6 mm, 5 µm column with a mobile phase composed of 10mM ammonium formate-methanol-acetonitrile in the ratio of 50:25:25 (v/v/v), at a flow rate of 1.0 mL/min. Turbo ion-spray interface (TIS) operated in positive ionization mode was used for the mass spectrometric detection. The MRM transitions monitored were m/z 230.1/112.1 (lamivudine), m/z 288.0/176.2 (tenofovir), m/z 316.2/168.1 (Efavirenz), m/z 248.1/130.1 (Emtricitabine,ISTD) and m/z 287.2/191.2(Abacavir,ISTD). The method was developed and the established calibration ranges are 10-3018 ng/mL, 5-500ng/mL and 20-6021ng/mLfor lamivudine, tenofovir disoproxilfumarateand efavirenz respectively. The slope values are consistent and regression values were found to be more than or equal to 0.99. The accuracy for run size reproducibility batch and dilution integrity (1/2, 1/4) for lamivudine, tenofovir disoproxilfumarateand efavirenz was found to be in the range of 96.9-102.2%. The precision was found to be less than 8.1% for three analytes. Further, the reported method was validated as per the ICH guidelines and found to be well with in the acceptable range. Results: The method showed good accuracy, low limits of quantification, adequate recovery, minimal matrix effects, and specificity. Analyte stability under multiple storage conditions was demonstrated. Conclusion: The validated LC-MS/MS method provides a reliable tool for therapeutic drug monitoring and pharmacokinetic studies of anti-HIV regimens. The assay can be applied to large populations, especially in resource-poor settings, to help individualize dosing and improve clinical outcomes while reducing toxicity

Authors

Publication Details

Journal
Journal of Neonatal Surgery
Published
2025-05-01
DOI
https://doi.org/10.63682/jns.v14i19s.4966
Primary Topic
HIV/AIDS drug development and treatment
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development and Validation of an LC-MS/MS Method for the Determination of Lamivudine, Tenofovir & Efavirenz in human plasma

A. Seetha Devi, Mounika V Mounika
1 links
Journal of Neonatal Surgery
HIV/AIDS drug development and treatment
article

Development and Validation of an LC-MS/MS Method for the Determination of Lamivudine, Tenofovir & Efavirenz in human plasma

A. Seetha Devi, Mounika V Mounika
article en

Abstract

Introduction: Therapeutic drug monitoring of Lamivudine, tenofovir and efavirenz, three commonly used antiretroviral drugs, is important to maximize effectiveness while minimizing side effects. Materials and Methods: Chromatographic separation was done on Xterra, C18 (2), 150 X 4.6 mm, 5 µm column with a mobile phase composed of 10mM ammonium formate-methanol-acetonitrile in the ratio of 50:25:25 (v/v/v), at a flow rate of 1.0 mL/min. Turbo ion-spray interface (TIS) operated in positive ionization mode was used for the mass spectrometric detection. The MRM transitions monitored were m/z 230.1/112.1 (lamivudine), m/z 288.0/176.2 (tenofovir), m/z 316.2/168.1 (Efavirenz), m/z 248.1/130.1 (Emtricitabine,ISTD) and m/z 287.2/191.2(Abacavir,ISTD). The method was developed and the established calibration ranges are 10-3018 ng/mL, 5-500ng/mL and 20-6021ng/mLfor lamivudine, tenofovir disoproxilfumarateand efavirenz respectively. The slope values are consistent and regression values were found to be more than or equal to 0.99. The accuracy for run size reproducibility batch and dilution integrity (1/2, 1/4) for lamivudine, tenofovir disoproxilfumarateand efavirenz was found to be in the range of 96.9-102.2%. The precision was found to be less than 8.1% for three analytes. Further, the reported method was validated as per the ICH guidelines and found to be well with in the acceptable range. Results: The method showed good accuracy, low limits of quantification, adequate recovery, minimal matrix effects, and specificity. Analyte stability under multiple storage conditions was demonstrated. Conclusion: The validated LC-MS/MS method provides a reliable tool for therapeutic drug monitoring and pharmacokinetic studies of anti-HIV regimens. The assay can be applied to large populations, especially in resource-poor settings, to help individualize dosing and improve clinical outcomes while reducing toxicity

Journal of Neonatal SurgeryVol. 14(19S)
Openalex Percentile: Top 88%
HIV/AIDS drug development and treatment
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Development and Validation of an LC-MS/MS Method for the Determination of Lamivudine, Tenofovir & Efavirenz in human plasma — A. Seetha Devi, Mounika V Mounika · Journal of Neonatal Surgery (2025) | TGRS Research Map | TGRS