Design and characterization of practical small-field collimators for radiotherapy calibration unit at a secondary standard dosimetry laboratory

We aimed to design small fields collimators made of lead blocks, with diameters of 40 mm, 30 mm, 20 mm, and 10 mm, for the calibration of small volume radiotherapy ionization chambers using the Algerian Secondary Standard Dosimetry Laboratory (SSDL) ELDORADO 78<sup>60</sup>Co calibration unit. The radiation properties of the designed collimators were assessed using GAFChromic EBT3 dosimetry film (Ashland, USA) and two PTW small volume ionization chambers (PTW, Freiburg, Germany). Additionally, the photons spectra for the four different fields were also investigated using Monte Carlo (MC) method. The MC method was also utilized to define the dimensions of the collimators that achieve the desired field sizes at the reference distance of 80 cm. We simulated the photon beam of the<sup>60</sup>Co source from the irradiation unit to estimate the influencing quantities including field size and beam attenuation. EBT3 film, PTW Pinpoint and PTW Semiflex ionization chambers were used to measure the beam profiles, the output factors (OF), and the percentage depth doses (PDD) for the designed collimators. These measurements were performed at a depth of 5 cm and a source-to-surface distance of 80 cm. The beam profiles simulated with MC method and experimentally measured were found to be agreed within 3%/1 mm gamma criteria. The ionization chambers underestimate the OFs in comparison to EBT3 film for the four field sizes. The PDD curves were measured by the Pinpoint chamber and calculated using MC simulation. Significant differences were noted between the curves of the 20 and 10 mm collimators compared to those of the 40 mm and 30 mm ones. The curves were in agreement with the relative simulated data. Custom small-field collimators were successfully produced and characterized, which will facilitate studies on the variation and stability of calibration coefficients for small volume detectors operating within small-fields at our SSDL.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2025-05-02
DOI
https://doi.org/10.1088/2057-1976/add34f
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Design and characterization of practical small-field collimators for radiotherapy calibration unit at a secondary standard dosimetry laboratory

Mehenna Arib, Toufik Medjadj, Silia Hayoune
Biomedical Physics & Engineering Express
Advanced Radiotherapy Techniques
article

Design and characterization of practical small-field collimators for radiotherapy calibration unit at a secondary standard dosimetry laboratory

Mehenna Arib, Toufik Medjadj, Silia Hayoune
article en

Abstract

We aimed to design small fields collimators made of lead blocks, with diameters of 40 mm, 30 mm, 20 mm, and 10 mm, for the calibration of small volume radiotherapy ionization chambers using the Algerian Secondary Standard Dosimetry Laboratory (SSDL) ELDORADO 78<sup>60</sup>Co calibration unit. The radiation properties of the designed collimators were assessed using GAFChromic EBT3 dosimetry film (Ashland, USA) and two PTW small volume ionization chambers (PTW, Freiburg, Germany). Additionally, the photons spectra for the four different fields were also investigated using Monte Carlo (MC) method. The MC method was also utilized to define the dimensions of the collimators that achieve the desired field sizes at the reference distance of 80 cm. We simulated the photon beam of the<sup>60</sup>Co source from the irradiation unit to estimate the influencing quantities including field size and beam attenuation. EBT3 film, PTW Pinpoint and PTW Semiflex ionization chambers were used to measure the beam profiles, the output factors (OF), and the percentage depth doses (PDD) for the designed collimators. These measurements were performed at a depth of 5 cm and a source-to-surface distance of 80 cm. The beam profiles simulated with MC method and experimentally measured were found to be agreed within 3%/1 mm gamma criteria. The ionization chambers underestimate the OFs in comparison to EBT3 film for the four field sizes. The PDD curves were measured by the Pinpoint chamber and calculated using MC simulation. Significant differences were noted between the curves of the 20 and 10 mm collimators compared to those of the 40 mm and 30 mm ones. The curves were in agreement with the relative simulated data. Custom small-field collimators were successfully produced and characterized, which will facilitate studies on the variation and stability of calibration coefficients for small volume detectors operating within small-fields at our SSDL.

Biomedical Physics & Engineering ExpressVol. 11(3)
King Faisal Specialist Hospital & Research Centre (SA), Centre de Recherche Nucléaire d’Alger (DZ)
Affordable and clean energy
Openalex Percentile: Top 89%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.