A coil-channel multiplexer for high-density receive arrays compatible with the MR-linac

<i>Objective.</i>We present a prototype add-on multiplexer for retro-fitting high-channel count receive coils onto systems with analog readout chains, such as the MR-linac. This particular system is currently equipped with two coil elements, containing only four channels each.<i>Approach.</i>We developed a 4:1 multiplexer board based on time-division. CMOS-based single-pole double-throw switches were used to accommodate for fast switching and adequate encoding bandwidths. Coil-channel images were separated after reconstruction by solving a system of equations based on the square-wave modulation imposed by the multiplexer. The prototype was tested at a 1.5 T scanner on an image-quality phantom and a slice of bovine muscle using four coil channels from a previously characterized 16-channel anterior coil prototype.<i>Main results.</i>The image data received through multiplexing could be reliably divided into separate coil-channel images, preserving the coil sensitivity maps. Leakage between channels remained below -30 dB. In each coil channel the signal-to-noise ratio was penalized by a factor of two compared to direct digitization due to the inflicted data compression. Furthermore, it was shown that the<i>g</i>-factor for parallel imaging remained largely unaffected by multiplexing. Imaging speedup is expected with this multiplexer, since a fourfold channel-count increase typically yields<i>g</i>-factor improvements that are superior to the signal-quality reduction.<i>Significance.</i>A 4:1 coil-channel add-on multiplexer was developed and its functionality demonstrated in the MR readout chain at 1.5 T. We expect that this solution will enable significant imaging speedup for clinical MR systems that suffer from compatibility with a limited number of channels for the receive coil array.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-05-02
DOI
https://doi.org/10.1088/1361-6560/add3b7
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A coil-channel multiplexer for high-density receive arrays compatible with the MR-linac

Cornelis A. T. van den Berg, Tom Bosma, Martin F. Fast, Pim Borman et al.
Physics in Medicine and Biology
Advanced MRI Techniques and Applications
article

A coil-channel multiplexer for high-density receive arrays compatible with the MR-linac

Cornelis A. T. van den Berg, Tom Bosma, Martin F. Fast, Pim Borman, J G M Kok
article en

Abstract

<i>Objective.</i>We present a prototype add-on multiplexer for retro-fitting high-channel count receive coils onto systems with analog readout chains, such as the MR-linac. This particular system is currently equipped with two coil elements, containing only four channels each.<i>Approach.</i>We developed a 4:1 multiplexer board based on time-division. CMOS-based single-pole double-throw switches were used to accommodate for fast switching and adequate encoding bandwidths. Coil-channel images were separated after reconstruction by solving a system of equations based on the square-wave modulation imposed by the multiplexer. The prototype was tested at a 1.5 T scanner on an image-quality phantom and a slice of bovine muscle using four coil channels from a previously characterized 16-channel anterior coil prototype.<i>Main results.</i>The image data received through multiplexing could be reliably divided into separate coil-channel images, preserving the coil sensitivity maps. Leakage between channels remained below -30 dB. In each coil channel the signal-to-noise ratio was penalized by a factor of two compared to direct digitization due to the inflicted data compression. Furthermore, it was shown that the<i>g</i>-factor for parallel imaging remained largely unaffected by multiplexing. Imaging speedup is expected with this multiplexer, since a fourfold channel-count increase typically yields<i>g</i>-factor improvements that are superior to the signal-quality reduction.<i>Significance.</i>A 4:1 coil-channel add-on multiplexer was developed and its functionality demonstrated in the MR readout chain at 1.5 T. We expect that this solution will enable significant imaging speedup for clinical MR systems that suffer from compatibility with a limited number of channels for the receive coil array.

Physics in Medicine and BiologyVol. 70(10)
Heidelberg University (DE), University Hospital Heidelberg (DE), University Medical Center Utrecht (NL)
Openalex Percentile: Top 86%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.