Overcoming methodological barriers in electronic nose clinical studies, a simulation data-based approach

Analysis of volatile organic compounds by electronic nose (e-nose) may address gaps in non-invasive screening for neoplasia. Machine learning impacts study design and sample size requirements, but guidance on clinical study design is limited. This study evaluates how neoplasia prevalence, augmented data, and the number of e-nose devices impact sample size requirements. Simulated e-nose breath test data were created using real-world study data. We examined the effect of varying neoplasia prevalence (50%-5%) and data augmentation on model performance, as well as the impact of using multiple devices. Prediction models were developed using single value decomposition and random forest, and convolutional neural networks. Model performance was displayed as area under the receiver operating characteristics curve and F1-score. Stable model performance was defined as the phase where additional data no longer increases model performance. We found that lower neoplasia prevalence significantly increased sample size requirements, with low-prevalence settings (5%) requiring up to five times more data than high-prevalence settings (50%) for stable model performance. Model performance varied between devices, and integrating data from multiple devices required larger sample sizes. Approximately 400 data points per device at 50% prevalence, and 2100 data points at 5% prevalence, were necessary to reach stable model performance. Concluding, sample size requirements for e-nose studies are heavily influenced by disease prevalence and the number of devices used. Limiting device variability and ensuring sufficient case and control samples per device are crucial for achieving reliable predictive performance. Specific requirements will vary based on sensor and disease characteristics.<b>ClinicalTrials.gov Identifier:</b>Clinicaltrials.gov Identifier NCT03346005 (model study) and NCT04357158 (validation study).

Authors

Institutions

Publication Details

Journal
Journal of Breath Research
Published
2025-05-01
DOI
https://doi.org/10.1088/1752-7163/add291
Citations
3
Primary Topic
Advanced Chemical Sensor Technologies
Type
article
Field-Weighted Citation Impact
2.02

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Overcoming methodological barriers in electronic nose clinical studies, a simulation data-based approach

Job A.M. van der Palen, Milou L. M. van Riswijk, Ruth Lucas, Peter D. Siersema et al.
3 citations
Journal of Breath Research
Advanced Chemical Sensor Technologies
2.02
article

Overcoming methodological barriers in electronic nose clinical studies, a simulation data-based approach

Job A.M. van der Palen, Milou L. M. van Riswijk, Ruth Lucas, Peter D. Siersema, Bastiaan van Tintelen
article en
3 citations

Abstract

Analysis of volatile organic compounds by electronic nose (e-nose) may address gaps in non-invasive screening for neoplasia. Machine learning impacts study design and sample size requirements, but guidance on clinical study design is limited. This study evaluates how neoplasia prevalence, augmented data, and the number of e-nose devices impact sample size requirements. Simulated e-nose breath test data were created using real-world study data. We examined the effect of varying neoplasia prevalence (50%-5%) and data augmentation on model performance, as well as the impact of using multiple devices. Prediction models were developed using single value decomposition and random forest, and convolutional neural networks. Model performance was displayed as area under the receiver operating characteristics curve and F1-score. Stable model performance was defined as the phase where additional data no longer increases model performance. We found that lower neoplasia prevalence significantly increased sample size requirements, with low-prevalence settings (5%) requiring up to five times more data than high-prevalence settings (50%) for stable model performance. Model performance varied between devices, and integrating data from multiple devices required larger sample sizes. Approximately 400 data points per device at 50% prevalence, and 2100 data points at 5% prevalence, were necessary to reach stable model performance. Concluding, sample size requirements for e-nose studies are heavily influenced by disease prevalence and the number of devices used. Limiting device variability and ensuring sufficient case and control samples per device are crucial for achieving reliable predictive performance. Specific requirements will vary based on sensor and disease characteristics.<b>ClinicalTrials.gov Identifier:</b>Clinicaltrials.gov Identifier NCT03346005 (model study) and NCT04357158 (validation study).

Journal of Breath ResearchVol. 19(3)
Radboud University Nijmegen (NL), Medisch Spectrum Twente (NL), Erasmus MC (NL), Radboud University Medical Center (NL), The eNose (Netherlands) (NL), Erasmus University Rotterdam (NL)
Maag Lever Darm Stichting
Life in Land
Openalex Percentile: Top 14%
Advanced Chemical Sensor Technologies
2.02
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.