Assessing the haemodynamics of the cerebral venous system anatomy in multiple sclerosis and a healthy control using in vivo and 3D printed in vitro modelling

Abstract This study aimed to compare computational fluid dynamics (CFD) results to those acquired in vivo with 4D Flow magnetic resonance imaging (MRI) and in vitro with a 3D printed model using pressure catheter manometry. The goal was to investigate the haemodynamics of the cerebral venous system (CVS) and assess the accuracy of the methodologies, to highlight any discrepancies between the techniques. One participant living with multiple sclerosis (MS) and one healthy control were recruited for this study. MRI was performed to generate 3D geometries of the anatomy and to compute blood flow rates at the boundaries, with 4D Flow MRI velocity streamlines for the control participant. CFD models were created for the two participants and simulated using the patient-specific boundary conditions. A 3D printed geometry of the MS participant was created and a flow loop experiment was conducted to measure the cerebral venous pressures. The venous pressures were found to be comparable to that observed in the CFD simulation. 4D Flow MRI velocity streamlines of the CVS were found to correspond well to the CFD findings, except for a few regions, which were likely impacted by the low resolution of the MRI. The use of all three methods enabled the successful validation of the velocity, flow features and pressure, and ensured that the haemodynamics of the CVS as resolved using CFD, were accurate. This highlights the potential for increased efficacy of the clinical outcomes of future studies that utilise such methods. Graphical abstract

Authors

Institutions

Publication Details

Journal
Experiments in Fluids
Published
2025-05-01
DOI
https://doi.org/10.1007/s00348-025-04028-7
Primary Topic
Acute Ischemic Stroke Management
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Assessing the haemodynamics of the cerebral venous system anatomy in multiple sclerosis and a healthy control using in vivo and 3D printed in vitro modelling

Alexander Robert Bateman, Tracie Barber, Jeannette Lechner‐Scott, Saadallah Ramadan et al.
1 links
Experiments in Fluids
Acute Ischemic Stroke Management
article

Assessing the haemodynamics of the cerebral venous system anatomy in multiple sclerosis and a healthy control using in vivo and 3D printed in vitro modelling

Alexander Robert Bateman, Tracie Barber, Jeannette Lechner‐Scott, Saadallah Ramadan, Pujith Rajaguru Senapathy Vijayaratnam, Grant A. Bateman, Shiami Luchow
article en

Abstract

Abstract This study aimed to compare computational fluid dynamics (CFD) results to those acquired in vivo with 4D Flow magnetic resonance imaging (MRI) and in vitro with a 3D printed model using pressure catheter manometry. The goal was to investigate the haemodynamics of the cerebral venous system (CVS) and assess the accuracy of the methodologies, to highlight any discrepancies between the techniques. One participant living with multiple sclerosis (MS) and one healthy control were recruited for this study. MRI was performed to generate 3D geometries of the anatomy and to compute blood flow rates at the boundaries, with 4D Flow MRI velocity streamlines for the control participant. CFD models were created for the two participants and simulated using the patient-specific boundary conditions. A 3D printed geometry of the MS participant was created and a flow loop experiment was conducted to measure the cerebral venous pressures. The venous pressures were found to be comparable to that observed in the CFD simulation. 4D Flow MRI velocity streamlines of the CVS were found to correspond well to the CFD findings, except for a few regions, which were likely impacted by the low resolution of the MRI. The use of all three methods enabled the successful validation of the velocity, flow features and pressure, and ensured that the haemodynamics of the CVS as resolved using CFD, were accurate. This highlights the potential for increased efficacy of the clinical outcomes of future studies that utilise such methods. Graphical abstract

Experiments in FluidsVol. 66(5)
John Hunter Hospital (AU), UNSW Sydney (AU), Hunter Medical Research Institute (AU), University of Newcastle Australia (AU)
Australian Government, National Imaging Facility, Hunter Medical Research Institute, University of New South Wales, Siemens Healthineers
Openalex Percentile: Top 87%
Acute Ischemic Stroke Management
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.