Portable turbidimetric device for in-time monitoring of bacterial contamination in drinking water

Microbial contamination of drinking water can affect public health with acute effects even after short-term consumption. The delays and limitations of current methods of drinking water quality control hinder rapid responses to unforeseen contamination incidents. Therefore, we designed and developed a compact and cost-effective device that can be integrated into the continuous processes of water treatment plants for indicative monitoring and in-time detection of bacterial contamination. The flow-through analyzer operates by real-time culturing bacteria and measuring turbidity, allowing the analysis of bacterial contamination with high sensitivity – detection of a single cell per 100 mL in a temperature-controlled environment was successfully demonstrated. Practical verification of the method efficacy was conducted using water samples obtained from a local water treatment facility. Furthermore, the analyzer was tested to assess the potential for enhancing selectivity by increasing the operating temperature from 37 °C to 44 °C. This allows more selective detection of E. coli , a model organism of fecal contamination, in contrast to other bacterial strains typically present at low concentrations in drinking water. • A cost-effective device for in-time detection of bacterial contamination in water • Capability of continuous on-site monitoring • Simple device construction and robust performance for pre-screening measurements • Comparison of device performance with reference methods • Enhanced E. coli detection selectivity by modulation of measurement temperature

Authors

Institutions

Publication Details

Journal
Journal of Water Process Engineering
Published
2025-05-01
DOI
https://doi.org/10.1016/j.jwpe.2025.107832
Citations
1
Primary Topic
Water Quality Monitoring Technologies
Type
article
Field-Weighted Citation Impact
0.82
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Portable turbidimetric device for in-time monitoring of bacterial contamination in drinking water

Zdeněk Farka, Petr Skládal, Karel Lacina, Radka Obořilová et al.
1 citations
Journal of Water Process Engineering
Water Quality Monitoring Technologies
0.82
article

Portable turbidimetric device for in-time monitoring of bacterial contamination in drinking water

Zdeněk Farka, Petr Skládal, Karel Lacina, Radka Obořilová, Jiří Buzík
article en
1 citations

Abstract

Microbial contamination of drinking water can affect public health with acute effects even after short-term consumption. The delays and limitations of current methods of drinking water quality control hinder rapid responses to unforeseen contamination incidents. Therefore, we designed and developed a compact and cost-effective device that can be integrated into the continuous processes of water treatment plants for indicative monitoring and in-time detection of bacterial contamination. The flow-through analyzer operates by real-time culturing bacteria and measuring turbidity, allowing the analysis of bacterial contamination with high sensitivity – detection of a single cell per 100 mL in a temperature-controlled environment was successfully demonstrated. Practical verification of the method efficacy was conducted using water samples obtained from a local water treatment facility. Furthermore, the analyzer was tested to assess the potential for enhancing selectivity by increasing the operating temperature from 37 °C to 44 °C. This allows more selective detection of E. coli , a model organism of fecal contamination, in contrast to other bacterial strains typically present at low concentrations in drinking water. • A cost-effective device for in-time detection of bacterial contamination in water • Capability of continuous on-site monitoring • Simple device construction and robust performance for pre-screening measurements • Comparison of device performance with reference methods • Enhanced E. coli detection selectivity by modulation of measurement temperature

Journal of Water Process EngineeringVol. 74
Central European Institute of Technology (CZ), Masaryk University (CZ), Central European Institute of Technology – Masaryk University (CZ)
Clean water and sanitation
Openalex Percentile: Top 27%
Water Quality Monitoring Technologies
0.82
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.